background image

 

 
For these reasons, it is recommended that a recalibration 
cycle  be  established  for  each  accelerometer.    This 
schedule  is  unique  and  is  based  on  a  variety  of  factors, 
such  as  extent  of  use,  environmental  conditions, 
accuracy requirements, trend information obtained from 
previous  calibration  records,  contractual  regulations, 
frequency of “cross-checking” against other equipment, 
manufacturer  recommendation,  and  any  risk  associated 
with incorrect readings.  International standards, such as 
ISO 10012-1, provide insight and suggested methods for 
determining  recalibration  intervals  for  most  measuring 
equipment.    With  the  above  information  in  mind  and 
under  “normal”  circumstances,  PCB  conservatively 
suggests  a 12- to 24-month recalibration cycle for most 
piezoelectric accelerometers.   
 

Note:

 

It  is  good  measurement  practice  to  verify  the 

performance  of  each  accelerometer  with  a  Handheld 
Shaker or other calibration device before and after each 
measurement.    The  PCB  Model  394C06  Handheld 
Shaker  operates  at  a  fixed  frequency  and  known 
amplitude  (1.0  g)  to  provide  a  quick  check  of  sensor 
sensitivity.

 

 

7.1 SENSOR RECALIBRATION 

Accelerometer  recalibration  services  are  typically 
performed  by  PCB’s  internal  metrology  laboratory.  
(Other  international  and  private  laboratories  are  also 
available.)  The PCB laboratory is certified to ISO 9001, 
accredited  by  A2LA  to  ISO  17025,  complies  with  ISO 
10012-1  (and  former  MIL-STD-45662A),  and  uses 
equipment directly traceable to N.I.S.T.  This assures an 
accurate calibration of relevant specifications. 
 
In  addition,  many  companies  choose  to  purchase  the 
equipment  necessary  to  perform  the  recalibration 
procedure  themselves.    While  this  may  result  in  both  a 
savings of time and money, it has also been attributed to 
incorrect  readings  and  costly  errors.    Therefore,  in  an 
effort  to  prevent  the  common  mistakes  associated  with 
customer-performed  calibration,  this  document  includes 
a  broad  overview  of  the  Back-to-Back  Calibration 
technique.    This  technique  provides  a  quick  and  easy 
method  for  determining  the  sensitivity  of  a  test 
accelerometer over a wide frequency range.   

 
7.2 BACK-TO-BACK CALIBRATION THEORY 

Back-to-Back  Calibration  is  perhaps  the  most  common 
method  for  determining  the  sensitivity  of  piezoelectric 
accelerometers.    This  method  relies  on  a  simple 
comparison  to  a  previously  calibrated  accelerometer, 
typically referred to as a reference standard.   
 
 

 

 

 

Figure 9.

 Reference Standard Accelerometer 

 
These  high-accuracy  devices,  which  are  directly 
traceable  to  a  recognized  standards  laboratory,  are 
designed  for  stability,  as  well  as  configured to accept a 
test accelerometer.  By mounting a test accelerometer to 
the  reference  standard  and  then  connecting  this 
combination to a suitable vibration source, it is possible 
to  vibrate  both  devices  and  compare  the  data  as  shown 
in Figure 10.  (Test set-ups may be automated and vary, 
depending  on  the  type  and  number  of  accelerometers 
being calibrated.) 
 

 

 
Figure 10.

 Typical Back-to-Back Calibration System 

 
Because  the  acceleration  is  the  same  on  both  sensors, 
the ratio of their outputs (V

T

/V

R

) must also be the ratio 

of  their  sensitivities.    With  the  sensitivity  of  the 
reference  standard  (S

R

)  known,  the  exact  sensitivity  of 

the  test  sensor  (S

T

)  is  easily  calculated  by  using  the 

following equation: 
 
 

 

S

= S

R

 (V

T

/V

R

 
By  varying  the  frequency  of  the  vibration,  the  sensor 
may  be  calibrated  over  its  entire  operating  frequency 
range.    The  typical  response  of  an  unfiltered 
accelerometer is shown in Figure 11. 

Summary of Contents for 357B54

Page 1: ...lerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mai...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Page 7: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Page 8: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Page 9: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Page 10: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Page 11: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Page 12: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Page 13: ...etic Size Hex x Height 3 4 in x 1 43 in 3 4 in x 36 3 mm Weight 1 80 oz 51 gm 1 Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Top Top Mounting Thread 10 32...

Page 14: ......

Page 15: ...ONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M...

Reviews: