Drawing Number: 21096 
Revision: NR

 

 

SERIES 134A20 TOURMALINE PRESSURE BAR GENERAL OPERATION MANUAL

 

Note  that  the  Series  402A  In-Line  Amplifier 
features  a  frequency  response  of  greater  than  1 
MHz. A charge amplifier used  with the tourmaline 
pressure  bar  may  limit  the  upper  frequency 
response of the system. 
 
Connecting  the  Series  134A  Tourmaline  Pressure 
Bar directly to a readout instrument without the use 
of  the  Series  402A  In-Line  Amplifier  or  other 
impedance-converting  device  can  be  done  with 
some  limitations.  The  input  impedance  of  the 
readout  instrument  may  be  significantly  low  to 
create  a  very  high-pass  filtering  effect.  This  can 
affect  measurement  of  pulse  duration  or  amplitude 
of a slow-rising pulse. Also, the voltage sensitivity 
of the output of the tourmaline bar into the readout 
instrument is affected by the cable capacitance and 
input capacitance of the readout instrument. 
 
Contact  the  factory  with  any  questions  regarding 
use of the sensor in the charge output mode. 
 

6.1 

USE OF THE IN-LINE 

 

AMPLIFIER

 

 
The  Series  402A  In-Line  Amplifier  is  used  to 
convert  the  high-impedance  output  of  the 
tourmaline  pressure  bar  into  a  low-impedance 
output. The in-line amplifier may also be used as a 
unity gain amplifier for checking system frequency 
response.  A  signal  generator  is  connected  to  the 
input  of  the  in-line  amplifier  and  fed  through  the 
system. 
 

7.0 

MAINTENANCE AND 

 

REPAIR

 

 
The  electrical  connectors  on  the  tourmaline 
pressure bar, the input of the in-line amplifier, and 
the  three-inch,  low-noise  cable  must  be  kept  clean 
and  dry,  especially  if  they  are operating in a dusty 
or  wet  environment.  This is to prevent drift due to 
low insulation resistance.  Also, note the following 
precautions: 
 
1. Do not exceed maximum levels. 
 

2. Constant temperatures around the tourmaline bar 
should be limited to 150

F. Short-term exposure to 

higher  temperatures,  such  as  that  found  in  a  blast 
wave, do not cause damage. 
 
3.  Use  only  PCB  power  conditioners  or  approved 
alternative. 
 
4.  Current  to  the  in-line  amplifier  must  be  limited 
to not more than 20 mA to avoid damage. 

5.  If  the  tourmaline  pressure  bar  and  in-line 
amplifier  are  to  be  left  outside  overnight  or  in 
humid or rainy environments, they should be coated 
with silicone oil and covered. 

Summary of Contents for 134A22

Page 1: ...the 402A02 installation and operating manuals that comprise a Model 134A22 Tourmaline ICP pressure bar 5000 psi 1 mV psi 0 2 microsecond rise time kit For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 6: ...ies 134A Tourmaline Pressure Bar senses and transduces a pressure shock wave into a high impedance electrostatic charge This charge is sent through the three inch low noise cable into the Series 402A In Line Amplifier where it is converted into a low impedance voltage with less than 100 ohms output impedance See Figure 2 The circuit in the in line amplifier consists of a MOSFET input IC with a ver...

Page 7: ...er conditioner or in the power conditioner Allow the system to stabilize for a minute A signal drift may occur when connecting the power conditioner to the readout instrument This should stabilize as components in the power conditioner and readout instrument charge If long cables between the Series 134A20 and the power conditioner are used impedance matching of the long cable and increase of the c...

Page 8: ...FIER The Series 402A In Line Amplifier is used to convert the high impedance output of the tourmaline pressure bar into a low impedance output The in line amplifier may also be used as a unity gain amplifier for checking system frequency response A signal generator is connected to the input of the in line amplifier and fed through the system 7 0 MAINTENANCE AND REPAIR The electrical connectors on ...

Page 9: ...ifier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 10: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 11: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 12: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 13: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 14: ......

Page 15: ......

Page 16: ......

Page 17: ......

Page 18: ...the 402A02 installation and operating manuals that comprise a Model 134A22 Tourmaline ICP pressure bar 5000 psi 1 mV psi 0 2 microsecond rise time kit For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 19: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 20: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 21: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 22: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 23: ...poxy Epoxy Sealing Epoxy Epoxy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 1 4 oz 39 gm 4 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right tochange specifications without notice ICP is a registeredtrademarkof PCB Piezotronics Inc M Metric Mount N Negative Output Polarity OPTIONAL VERSION...

Page 24: ...ifier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 25: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 26: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 27: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 28: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 29: ......

Page 30: ...the 402A02 installation and operating manuals that comprise a Model 134A22 Tourmaline ICP pressure bar 5000 psi 1 mV psi 0 2 microsecond rise time kit For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 31: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Page 32: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Page 33: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Page 34: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Page 35: ......

Reviews: