background image

RECOMMENDED MAINTENANCE  

CONTINUED

Mantles.  Use Paulin mantles which are designed for use with propane and butane gas only.  Never use broken
mantles.  Inspect for carbon deposits each time gaslight is used.  Clean or replace defective mantles immediately.
Use proper ceramic burner nose and mantle combination.  (Example: “Tie-On” type burner nose, #L12-11A, with
tie-on mantles, #L13-6.  Or “Pre-formed” type burner nose, #L12-2A, with pre-formed mantles, #L13-2.)  Also see
“Mantle Installation and Lighting Instructions.”

Burner Nose Screen.  Screen must be in threaded end of burner nose.  Gaslight will not function properly if screen is
missing.

Seal.  Seal is placed on threaded end of burner nose.  Seal and burner nose should be screwed securely into bunsen.
The seal must be replaced anytime the burner nose is removed after the lamp has been used.

Bunsen.  Flow of combustion air and ventilation air must not be obstructed.  To ensure proper air flow and gas mixture,
the inside of the bunsen must be clean and free of any material that may potentially accumulate, such as spider webs,
insects, etc.  Make a periodic visual check of the Bunsen.  The Bunsen can be cleaned by drawing a clean, dry cloth
through it.  It can also be cleaned by blowing it out with compressed air.  Never attach a mantle directly to the Bunsen.

 

BUNSEN CLEANING INSTRUCTIONS

1.  Remove burner nose.

2.  Pull a clean, dry cloth
     through Bunsen to clean
    out foreign material.

3.  Replace seal.
     Reassemble burner
     nose and tighten securely.

     Note: Always replace seal if
     lamp has been previously lit.

or L13-8 Quick-Clip

Summary of Contents for Model 9

Page 1: ...r your gaslight unless you have been thoroughly trained and are experienced in such matters If your gaslight appears to be mal functioning turn it to the OFF position and immediately contact your local dealer or the Paulin factory After installation and before use have qualified gas personnel approve your installation and check for gas leaks using a leak detector Tubing fittings gaslight valve and...

Page 2: ...e the gaslight to malfunction Copper gas tubing Copper gas tubing must be internally tinned to meet National Gas Code Standards unless otherwise directed by local codes Nozzle All gaslights are equipped with standard propane nozzles To use gaslights at eleva tions of 4500 feet or more above sea level order High Altitude Nozzle L9 52A for propane gas or L9 61A for butane gas Note Never touch drille...

Page 3: ... Seal and burner nose should be screwed securely into bunsen The seal must be replaced anytime the burner nose is removed after the lamp has been used Bunsen Flow of combustion air and ventilation air must not be obstructed To ensure proper air flow and gas mixture the inside of the bunsen must be clean and free of any material that may potentially accumulate such as spider webs insects etc Make a...

Page 4: ...ubing 3 8 inch outside diameter is recommended Copper tubing must be internally tinned to meet National Fuel Gas Code unless otherwise directed by local codes Fittings If 3 8 inch copper tubing is used a 3 8 inch Flare Male Connector x 1 8 NPT Straight Fitting is required Paulin L22 12A included If Gaslight is to be connected from behind through wall a 1 8 NPT Street Elbow Male Female Paulin L22 2...

Page 5: ...ge 4 It is easier to connect the tubing before fastening Wall Bracket assembly to wall also Wall Bracket Assembly is less likely to be bent Valve Assembly D must be straight on the Wall Bracket Nozzle N must aim into the center of the Bunsen S for proper air and gas mixture 3 Fasten Wall Bracket Assembly to wall using four mounting screws A supplied CAUTION DO NOT OVERTIGHTEN SCREWS The four Bosse...

Page 6: ...opening so that it will fit over the ceramic burner nose and seat in lower groove of burner nose Do not tie the mantle to the Bunsen Distribute puckers in the mantle evenly around the burner nose Pull the ends of the string snugly so that mantle and string are seated securely in the burner nose groove Tie a double knot and clip off excess string 3 Replace globe 4 Mantle burn off WITH GASLIGHT VALV...

Page 7: ......

Page 8: ...ure to give full information including date and place of purchase full name and address 3 Manufacturer reserves the right to and at its option may refund the purchase price in full satisfaction of any warranty obligations hereunder 4 Manufacturer shall not be liable for incidental or consequential damages and in all events its liability hereunder shall not exceed the purchase price of the unit 5 M...

Reviews: