Operating manual of data acquisition system PWBlogg

The data logger PWBlogg N6740 is in accordance with the directive 

94/9EG (ATEX)

 as operatio-

nal supplements falling in category

 II 3 G

 for safe use within gas explosion hazard area of zone 2 

deliverable.

The data logger with ignition protection type nA is a non-sparking operating resource with low pow-

er input according to 

DIN

 

EN 60 079-15

.

Please, pay attention to the following note on safety:

• 

The online-outlet must only be used when there is no existence of an explosive atmosphere. 

When the system is not used, the online-outlet has to be closed using the protective cap.

• 

Check whether there is a necessity of embedding into the potential equalization of the equipment 

or the metering point. This potential equalization can happen via process (pressure) connection.

• 

Charging the accumulator must only be done outside of the ex-area (provided that there is no 

explosive atmosphere.

• 

Damaged devices must not be operated anymore. Please contact manufacturer for purpose of 

repair. 

•  All subsequent modifications to the device result in the consequence that a secure use

 

is no more guaranteed within ex-zone 2 and the EC-declaration of conformity loses ist 

validity.

3

Safety precautions

Power supply of the data measurement and storage device is provided by a built-in lithium ion ac-

cumulator. Please consider the following notes on safety during operation of the devices:

• 

Charging must only be done by the provided charger

•  Charging must be done away from easily inflammable items
• 

Do not short circuit lithium ion accumulator 

• 

Burning accumulators must never be extinguished by water! 

EXPLOSION HAZARD!

•  Do not throw into fire! 

EXPLOSION HAZARD!

• 

Do not charge unattended

• 

Do never open battery cell! 

EXPLOSION HAZARD!

• 

Do not use damaged accumulators!

• 

If electrolyte touches skin, immediately rinse skin with plenty of water and soap and consult a 

doctor. 

• 

If electrolyte touches eyes, immediately rinse the eyes with plenty of water and consult an oph-

thalmologist.

• 

Recycle completely discharged accumulators only.

Attention:  The  device  consists  of  electrostatically  sensitive  components.  Opening 

the device must only be done by manufacturer or by authorized persons.

Safety instructions for use in ex-zone 2

Summary of Contents for PWBlogg N6/40

Page 1: ...Operating manual Data acquisition system PWBlogg Type N6 40...

Page 2: ...llenstedt All rights reserved No section of the operating manual may be replicated or processed duplica ted or distributed using electronic systems in any way print photocopy microfilm or other me tho...

Page 3: ...repair All subsequent modifications to the device result in the consequence that a secure use is no more guaranteed within ex zone 2 and the EC declaration of conformity loses ist validity 3 Safety p...

Page 4: ...ters 6 2 2 Error sources 7 2 3 Data communication 7 3 Charging the built in accumulator 9 4 Technical specifications 9 4 1 Data logger 9 4 2 Explosion protection optional 10 4 3 Data memory 10 4 4 Est...

Page 5: ...r The measurement range begins at 50 mbar Data logger The data logger has an integrated analog digital converter with high resolution which is responsible for logging and digitizing pressure and tempe...

Page 6: ...resholds and records threshold violations by means of a se cond sample rate Additional to the threshold a hysteresis can be specified Important parameters to monitor thresholds are described as follow...

Page 7: ...s are non linearities of AD converter which are irrelevant on resolution of 8 bit or 12 bit However another important one is the digital step at small resolutions At the limit of two sequent digital v...

Page 8: ...8 Paul Wegener GmbH Performance data evaluation can be done in tabular form or as chart Individual periods can be enlarged by the integrated zoom function...

Page 9: ...in power saving mode If the accumulator s residual capacity falls below 20 then at the latest it should be recharged We recommend 100 charging prior to any operation The accumulator must not be replac...

Page 10: ...umber of channels occupied from preset sample intervals 1 and 2 as well as from storage and sample mode If the reading resolution is 8 bit 12 14 or 16 bit possible storage times will halve themselves...

Page 11: ...business and trade as well as small firms EN 61000 6 2 subject standard for interference resistance industrial sector EN 60079 0 electrical equipment for explosion hazard areas section 0 general requi...

Page 12: ...umulator is low Have it replaced by manufacturer Password forgotten Read out performance data send reading file to manufacturer Software does not read per formance data comple tely or reports errors w...

Page 13: ...Operating manual of data acquisition system PWBlogg 13...

Page 14: ...14 Paul Wegener GmbH...

Page 15: ...Operating manual of data acquisition system PWBlogg 15...

Page 16: ...r GmbH Tel Fax Internet e mail 49 0 39483 96 300 49 0 39483 96 400 www paul wegener de info paul wegener de Most recent emendment 01 08 2013 PWB FORM 2085 EN13 Paul Wegener GmbH Marienstra e 24 D 0649...

Reviews: