Introduction

 TD-8596A

3

012-08420B

Procedure

Set the plunger for a volume of 40 cc. 

Hold the base of the syringe firmly against a sturdy horizontal surface.

Slap down on the plunger with the palm of your hand to fully compress the 
gas inside the syringe. Hold this position until the temperature and pressure 
have equalized and are no longer changing (about 30 seconds). 

Release the plunger and allow it to return on its own. (It may not go back to 
40 cc.)

NOTE

: Do not use a hammer or mallet on the plunger! Only use the palm of 

your hand to push the plunger down.

Software Help

See the SPARKvue Help or PASCO Capstone Help for information about 
collecting, displaying, and analyzing data. 

In SPARKvue, select the HELP button (

) in any screen including the 

Home Screen.

In PASCO Capstone, select 

PASCO Capstone Help

 from the 

Help

 menu, or press 

F1

.

Capstone Experiment File

The included 

CAPSTONE Experiments Download Reference Sheet

 (012-15105) describes how to download pre-

configured PASCO Capstone experiments from the PASCO Web site for use with the TD-8596A Ideal Gas Law Apparatus. 

For TD-8596A Ideal Gas Law Apparatus:

To download the PASCO Capstone experiment, go to https://www.pasco.com/capstoneexperiments/ and download the files 
for EX-5527.

Double-click the downloaded “.ZIP” file to open the “EX-5527 Ideal Gas Law” folder.

The folder contains 
Capstone Files, Images, 
and a Word File.

Use PASCO Capstone to 
open the pre-configured 
Capstone workbook file.

Summary of Contents for TD-8596A

Page 1: ...com support pasco com 1 Not shown Included Equipment Included Equipment 1 Mechanical stop 4 Pressure connector quick release connector 2 Syringe and plunger 5 Temperature connector mini stereo jack 3...

Page 2: ...ected during the experiment to allow for different initial plunger positions NOTE If you are using a PASCO Wireless Pressure Sensor PS 3203 make a short adapter tube The Wireless Pressure Sensor inclu...

Page 3: ...or information about collecting displaying and analyzing data In SPARKvue select the HELP button in any screen including the Home Screen In PASCO Capstone select PASCO Capstone Help from the Help menu...

Page 4: ...ature and pressure have equalized and are no longer changing Record the final volume reading on the syringe 6 Stop recording data Analysis Constant Temperature 1 Highlight an area click and drag on th...

Page 5: ...V1 including your calculated value of Vo V1 is 40 cc 3 Highlight the area on the temperature graph where it peaks Pick the place where the temperature has peaked not the pressure It takes the temperat...

Page 6: ...SCO Capstone PASPORT and SPARKvue are trademarks or registered trademarks of PASCO scientific in the United States and or in other countries All other brands products or service names are or may be tr...

Page 7: ...ferent lab groups or over several trials you may find that calculated values of V0 vary by 1 cc or more This may look like a very large uncertainty but since it is added to the total volume of gas 40...

Page 8: ...eat flow from the environment into the syringe Experiment 1 Teacher s Notes Ideal Gas Law Electronic Workbook With the Capstone electronic workbook students will explore the relationship between the v...

Page 9: ...kPa K 300 K 102 kPa 62 8 cc 2 This figure is the volume of the syringe plus the volume of the attached tubing Volume of tubing 62 8 cc 60 0 cc 2 8 cc 3 The y intercept of the best fit line 2 68 cc is...

Page 10: ...Ideal Gas Law Apparatus TD 8596A 10 012 08420B 3 In theory the y intercepts of both slopes are equal because they both represent the volume of the tubing In this case they differ by 0 35 cc...

Reviews: