Chloride ISE Probe

Introduction

2

012-15721A

Introduction

This Chloride Ion Selective Electrode (ISE) Probe is a 

combination electrode that includes both reference and 

sensing half cells in one body housing. It is used for 

measuring chloride ion (Cl

-

) activity in aqueous samples. 

Technology: Replaceable Polymer/PVC Membrane 

(Chloride ISE Replacement Module PS-3550)

Overview

Required Equipment

PASCO Wireless pH/ISE/ORP Sensor (PS-3204)

or 

PASPORT High Precision pH/Temperature with 

ISE/ORP Amplifier (PS-2107) with a PASCO 

interface (see www.pasco.com for information)

or

Ion meter or pH/mV meter

PASCO data collection software (see 

www.pasco.com)

Wash bottle filled with distilled or de-ionized water 

Lab wipes 

Pipettes: 1 mL, 10 mL and 100 mL

Clean beakers*

*SE-7287 100 ml Beakers (12 pack) or SE-7288 1000 ml 

Beakers (6 pack)

Included Required Solutions*

Reference Fill Solution

1000 ppm Chloride Standard Solution

Ionic Strength Adjuster (ISE)

(*SC-3534 Chloride ISE Solutions)

Electrode Preparation

Replace the Module Blank

This electrode performs better when the reference 

junction is kept wetted during transit and storage. 

However, the sensor module is best stored and 

transported dry. Therefore, the electrode is shipped with 

a plastic Module Blank inside that must be removed prior 

to use. 

1. Unscrew the cap of the storage solution bottle (also 

known as “Soaker Bottle”) and remove the 

electrode. Unscrew the plastic Module Blank from 

the end of the electrode. Replace the blank with the 

included Calcium ISE Replacement Module. 

The Module Blank can be retained for long term 

storage. 

Caution: Do not touch the PVC sensor membrane 

on the end of the Calcium ISE Replacement Module 

with your fingers.

2. Thoroughly rinse the electrode with distilled water. 

Wipe carefully with a clean lab wipe.

3. Return the electrode to the storage bottle when it is 

not in use.

Fill the Electrode with Electrolyte

The reference (outer) chamber must be filled with 

Reference Fill Solution and remain open during testing:

Size and Material

Body (Epoxy): 12 mm OD x 155 

mm Length

Cap (ABS): 16 mm OD x 57 mm 

Length

Electrode Resistance

Less than one megaohm

Concentration Range

1.8 to 35,500 ppm

Reproducibility

±2%

pH Range

2 to 12

Temperature Range

0 to 80°C continuous:

80 to 100°C intermittent

Interferences

CN

-

, Br

-

, I

-

, OH

-

, S

2-

 must be 

absent, NH

3