Calibration

ation

The Wireless Optical Dissolved Oxygen Sensor is calibrated at the
factory and does not need to be recalibrated under most
circumstances. If needed, the sensor can be calibrated using one or
two known standards.

Pr

Prepar

epare the calibr

e the calibration standar

ation standards

ds

Prepare calibration standards that contain 0% and 100% dissolved
oxygen saturation. Either water-saturated air (recommended) or air-
saturated water can be used for the 100% dissolved oxygen
saturation standard.
• 100% dissolved oxygen saturation standard (water-saturated air)

100% dissolved oxygen saturation standard (water-saturated air)
Moisten the sponge inside the end of the rubber boot with water.
Loosely cover the sensor cap with the rubber boot. Let the sensor
acclimate for about 15 minutes. Leave the boot on during
calibration.

• 100% dissolved oxygen saturation standard (air-saturated water)

100% dissolved oxygen saturation standard (air-saturated water)
Connect an air stone to an air pump and put the air stone into a
beaker of water. Let the air pump to run for about 10 minutes.

• 0% dissolved oxygen saturation standard

0% dissolved oxygen saturation standard
Pour 150 milliliters of distilled water in a beaker. Dissolve sodium
sulfite, sodium bisulfite, or potassium sulfite until a small amount
remains undissolved in the bottom of a beaker.

Calibr

Calibrate the sensor using one standar

ate the sensor using one standardd

Use a 100% dissolved oxygen saturation standard to calibrate the
dissolved oxygen sensor.

SP

SPARK

ARKvue

vue

1. Click

.

2. Click

for DO

DO

22

Saturation

Saturation.

3. Under Calibration Type

Calibration Type, select 1 point (Adjust Slope Only)

1 point (Adjust Slope Only) then

click Continue

Continue.

4. Put the sensor in a 100% dissolved oxygen saturation standard.
5. Under Standard Value

Standard Value enter

100

then click Set Calibration

Set Calibration.

6. Click OK

OK.

P

PASCO Capst

ASCO Capstone

one

1. In the Tools

Tools palette, click Calibration

Calibration

.

2. Select Saturation

Saturation then click Next

Next.

3. Select One Standard (1 point slope)

One Standard (1 point slope) then click Next

Next.

4. Put the metal sensor body in a 100% dissolved oxygen saturation

standard.

5. Under Standard Value

Standard Value, enter

100

then click Set Current Value to

Set Current Value to

Standard Value

Standard Value.

6. Click Finish

Finish.

Calibr

Calibrate the sensor using two standar

ate the sensor using two standards

ds

Use a 0% and 100% dissolved oxygen saturation standards to
calibrate the dissolved oxygen sensor.

SP

SPARK

ARKvue

vue

1. Click

.

2. Click

for DO

DO

22

Saturation

Saturation.

3. Under Calibration Type

Calibration Type, select 2 point (Adjust Slope and Offset)

2 point (Adjust Slope and Offset)

then click Continue

Continue.

4. Put the metal sensor body in a 0% dissolved oxygen saturation

standard.

5. Under Calibration Point 1

Calibration Point 1, enter

0

for the Standard Value

Standard Value then

click Set Calibration

Set Calibration.

6. Put the sensor in a 100% dissolved oxygen saturation standard.
7. Under Standard Value

Standard Value enter

100

then click Set Calibration

Set Calibration.

8. Click OK

OK.

P

PASCO Capst

ASCO Capstone

one

1. In the Tools

Tools palette, click Calibration

Calibration

.

2. Select Saturation

Saturation then click Next

Next.

3. Select Two Standards

Two Standards then click Next

Next.

4. Put the metal sensor body in a 0% dissolved oxygen saturation

standard.

5. Under Standard Value

Standard Value, enter

0

then click Set Current Value to

Set Current Value to

Standard Value

Standard Value.

6. Put the metal sensor body in a 100% dissolved oxygen saturation

standard.

7. Under Standard Value

Standard Value, enter

100

then click Set Current Value to

Set Current Value to

Standard Value

Standard Value.

8. Click Finish

Finish.

Set the salinity constant

Set the salinity constant

The salinity of water can have an effect on the dissolved oxygen
measurement. A value can be entered for the salinity of the water
sample to account for this effect.

Note:

Note: It is recommended to use the PASPort Salinity Sensor

(PS-2195) to obtain a precise salinity measurement.

SP

SPARK

ARKvue

vue

1. Click

.

2. Click

for the Wireless Optical Dissolved Oxygen

Wireless Optical Dissolved Oxygen sensor.

3. Under Salinity

Salinity, enter a value in ppt.

P

PASCO Capst

ASCO Capstone

one

1. Click Hardware Setup

Hardware Setup

.

2. Click

for the Wireless Optical Dissolved Oxygen sensor.

3. Under Salinity

Salinity, enter a value in ppt.

4 | Wireless Optical Dissolved Oxygen Sensor