Using the sensor

Using the sensor

TTak

ake measur

e measurements in the lab

ements in the lab

SP

SPARK

ARKvue

vue

1. Turn on the sensor.

2. Open SPARKvue then click Sensor Data

Sensor Data

.

3. Select the Optical DO

Optical DO that matches the device ID.

4. Select DO

DO

22

Concentration

Concentration.

5. Select a Template

Template to display the data.

6. Remove the rubber boot to uncover the sensor cap.
7. Place the sensor into a water sample with the metal probe body

fully submerged.

8. Click Start

Start

to begin data collection.

P

PASCO Capst

ASCO Capstone

one

1. Turn on the sensor.
2. Click Sensor Data

Sensor Data

.

3. Select the Optical DO

Optical DO that matches the device ID.

4. Click Hardware Setup

Hardware Setup

to close it.

5. Remove the rubber boot to uncover the sensor cap.
6. Place the sensor into a water sample with the metal probe body

fully submerged.

7. Click Record

Record

to begin data collection.

TTak

ake measur

e measurements in the field

ements in the field

Use the sensor to take measurements in a stream or pond. Remote
logging is required when using the sensor in the field since the
wireless signal cannot be transmitted under water.

Note:

Note: The Metal Guard accessory (PS-3604) is required when

taking field measurements to protect the sensor cap. The accessory
includes a rope that attaches to the eye hook on the protective
cover.

Step 1: Set up r

Step 1: Set up remote logging

emote logging

SP

SPARK

ARKvue

vue

1. Turn on the sensor.

2. Open SPARKvue then click Remote Logging

Remote Logging

.

3. Select the Optical DO

Optical DO that matches the device ID.

4. Set the Sample Rate

Sample Rate using the left and right arrows.

Tip:

Tip: Decrease the sample rate to increase the logging time.

5. Click OK

OK and close SPARKvue.

P

PASCO Capst

ASCO Capstone

one

1. Turn on the sensor.
2. Open Capstone then click Remote Logging

Remote Logging

.

3. Select the Optical DO

Optical DO that matches the device ID.

4. Set the Sample Rate

Sample Rate using the up and down arrows.

Tip:

Tip: Decrease the sample rate to increase the logging time.

5. Click Start

Start and close Capstone.

Step 2: Collect the data

Step 2: Collect the data

1. Remove the rubber boot to uncover the sensor cap.
2. Install the metal guard on the probe.
3. Install the protective cover on the sensor body.

4. Attach the rope to the eye hook.
5. Submerge the sensor in a body of water to take measurements.
6. Turn off the sensor to stop data logging.

Step 3: Download and displa

Step 3: Download and display r

y remotely logged data

emotely logged data

SP

SPARK

ARKvue

vue

1. Turn on the sensor.

2. Open SPARKvue then click Remote Logging

Remote Logging

.

3. Select the Optical DO

Optical DO that matches the device ID.

4. Click Download Data

Download Data.

5. Select a Template

Template to display the data.

P

PASCO Capst

ASCO Capstone

one

1. Turn on the sensor.

2. Open Capstone then click Remote Logging

Remote Logging

.

3. Select the Optical DO

Optical DO that matches the device ID.

4. Click Download Data

Download Data.

5. Select Current experiment

Current experiment then click Download Data

Download Data.

6. Drag the Graph

Graph

display to the page.

7. Click Select Measurement

Select Measurement on the y-axis then select DO

DO

22

Concentration

Concentration from the menu.

Measur

Measure o

e oxy

xygen gas concentr

gen gas concentration in air

ation in air

The Wireless Optical Dissolved Oxygen Sensor was specifically
designed to be used in water. However, reasonably good qualitative
results may be obtained in air.
Oxygen gas concentration in air is calculated by taking the DO

2

Saturation value from the sensor and multiplying it by 20.9%. This is
the assumed partial pressure contribution of O

2

in air. Best accuracy

will be achieved in high humidity environments. Long term
measurements in dry air are not recommended as this will dry out
the probe sensor cap.

Note:

Note: The probe should not be used in an environment with

flammable, caustic, or corrosive gases.

Follow the same instructions as using the sensor in water, except
select O

O

22

Gas Concentration

Gas Concentration as the measurement.