background image

®

Technical Support

 PS

-

3201

7

013-14647C

Theory of Calibration

One of the functions of the PASCO Data Collection 

Software is to take the stream of raw data from a sensor 

and transform it into the calibrated data that you see in 

the Graph, Table, and other displays. If you do not 

calibrate a sensor yourself, the software uses a default 

calibration that is loaded when the sensor is connected.

You can think of the software as taking in raw data and 

outputing calibrated data. When you perform a 

calibration, the software redefines the linear equation 

that transforms the raw input data into the calibrated 

output data. The linear function is of the form:

Raw Input = Slope x Calibrated  Offset

Or:

Calibrated Output = (Raw Input - Offset)/Slope

The function can be represented graphically as a line.

Two points, Pt 1 and Pt 2, define the line. In the 

two-point calibration procedure, each point is reset by 

associating a known standard value (for instance, the 

temperature of ice water) with a raw input measurement 

that the sensor sends out when it is in that standard. In a 

one-point calibration, only one of the points is reset by 

the user.

Types of Calibration

There are three types of calibration: two-point, one-point 

slope, and one-point offset. Any of these calibrations can 

be performed on a single sensor, or simultaneously on 

multiple similar sensors; however, for any given sensor, 

the software will automatically select the most typical 

calibration type as the default setting.

Two-Point

In a two-point calibration, you reset two points to define 

a new line. This type of calibration affects both the slope 

and the offset. 

One-Point Slope

In a one-point slope calibration, you reset only one point. 

The slope of the line changes so that the line intersects 

the new point, while the offset (or Y-intercept) does not 

change. 

One-Point Offset

In a one-point offset calibration, you reset only one point. 

The line shifts so that it intersects the new point, but its 

slope does not change.

Offset calibration is usually used to make one sensor 

agree with another sensor. Due to normal variation 

among probes, a second probe might read consistently 

higher than the first probe. Normally this difference 

would be insignificant; however, an offset calibration can 

be used to bring the sensors into closer alignment.

Summary of Contents for PS-3201

Page 1: ...ter may cause a loss of wireless connection Do not immerse the sensor housing in boiling hot water Put only the Temperature Probe into the substance being measured ON OFF Information To turn the senso...

Page 2: ...Wireless Devices list that opens select the desired sensor that matches the XXX XXX Device ID number on the sensor Select Done Collect Data In SPARKvue select a measurement from the list under the sen...

Page 3: ...left to right until the indicator on the door is aligned with the second mark on the sensor Turn the sensor bottom side down so that the Battery Compartment Door can drop into the palm of your hand Ma...

Page 4: ...s Class A digital device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any...

Page 5: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Page 6: ...nsor is transferred to the Sensor Value box 10 Move the thermometer to the ice water bath Dry the temperature probe and then put the probe into the ice water bath with the thermometer 11 Wait until th...

Page 7: ...that standard In a one point calibration only one of the points is reset by the user Types of Calibration There are three types of calibration two point one point slope and one point offset Any of th...

Page 8: ...Wireless Temperature Sensor PS 3201 8 013 14647C...

Reviews: