background image

®

Wireless Temperature Sensor

PS

-

3201

6

013-14647C

Appendix B: Calibration

Calibration is not always necessary, especially if you are 

measuring a change in temperature rather than absolute 

temperature values. However, it is possible to calibrate 

the sensor. 

Prepare for Calibration

Calibration will need an ice-water bath, a container of 

hot water, and a thermometer. The sensor will need to 

be “paired” with a tablet or computer, and the data 

collection software (for example, SPARKvue or 

Capstone) should be running.

Using SPARKvue Software for Calibration

NOTE: Check the Help System in SPARKvue for the 

most up-to-date information.

1. Place the thermometer in the container of hot water.

2. Click (or press) the Experiment Tools button ( ).

The Experiment Tools screen opens.

3. Click Calibrate Sensor.

The Calibrate Sensor: Select Measurement screen 

opens.

4. Click the Sensor box and click the sensor to be 

calibrated.

5. Click the Calibration Type box and click a calibration 

type. (For this example, click “2-point”.)

6. Click Next.

The Calibrate Sensor Enter Values screen opens.

7. Place the temperature probe in the container of hot 

water.

8. Read the temperature on the thermometer. Under 

Calibration Point 1, click the Standard Value box 

and enter the known temperature value.

9. Under Calibration Point 1, click Read From Sensor.

The value measured by the sensor is transferred to 

the Sensor Value box.

10. Move the thermometer to the ice-water bath. Dry the 

temperature probe and then put the probe into the 

ice-water bath with the thermometer.

11. Wait until the reading on the thermometer stabilizes. 

Under Calibration Point 2, click the Standard Value 

box and enter the thermometer’s temperature value.

12. Under Calibration Point 2, click the Read From 

Sensor box.

The second value measured by the sensor is 

transferred to the Sensor Value box.

13. Click OK.

Using PASCO Capstone for Calibration

NOTE: Check the Help System in PASCO Capstone for 

the most up-to-date information.

1. Place the thermometer in the container of hot water.

2. Click Calibration ( ) in the Tools Palette.

3. Choose the measurement you would like to calibrate 

now: Temperature Measurement.

4. Click Next.

5. Choose the type of calibration you would like to 

perform: Two Standards (2 point).

6. Click Next.

7. Read the temperature on the thermometer and enter 

the value in the Standard Value text box.

8. Place the Temperature Probe in the container of hot 

water.

9. After the Current Value stabilizes, lick Set Current 

Value to Standard Value.

10. Click Next.

11. Move the thermometer to the ice-water bath. Dry the 

end of the temperature probe. Place the probe in the 

ice-water bath with the thermometer, 

12. Wait until the reading on the thermometer stabilizes. 

Enter the second known temperature value in the 

Standard Value text box.

13. After the Current Value stabilizes, click Set Current 

Value to Standard Value.

14. Click Next.

15. Click Finish. 

Summary of Contents for PS-3201

Page 1: ...ter may cause a loss of wireless connection Do not immerse the sensor housing in boiling hot water Put only the Temperature Probe into the substance being measured ON OFF Information To turn the senso...

Page 2: ...Wireless Devices list that opens select the desired sensor that matches the XXX XXX Device ID number on the sensor Select Done Collect Data In SPARKvue select a measurement from the list under the sen...

Page 3: ...left to right until the indicator on the door is aligned with the second mark on the sensor Turn the sensor bottom side down so that the Battery Compartment Door can drop into the palm of your hand Ma...

Page 4: ...s Class A digital device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any...

Page 5: ...hing lower than 0x6 means an older version of Bluetooth Your device will need the PS 3500 USB Bluetooth 4 0 Adapter 1The Mac Mini and MacBook Air were updated with Bluetooth SMART support in 2011 The...

Page 6: ...nsor is transferred to the Sensor Value box 10 Move the thermometer to the ice water bath Dry the temperature probe and then put the probe into the ice water bath with the thermometer 11 Wait until th...

Page 7: ...that standard In a one point calibration only one of the points is reset by the user Types of Calibration There are three types of calibration two point one point slope and one point offset Any of th...

Page 8: ...Wireless Temperature Sensor PS 3201 8 013 14647C...

Reviews: