background image

Instruction Manual

No. 012-08856A

Spirometer

PS-2152

Introduction

The PASPORT Spirometer allows students to collect accurate air-flow data from a pulmonary 
function test (PFT). The subject breathes through a sanitary, single-use mouthpiece as data is 
recorded and displayed on your computer or PASPORT interface. From the graphed data 
students can measure tidal volume, forced expiratory volume, and other important lung 
volumes and capacities.

Included Equipment

Part Number

Electronics enclosure and handle

PS-2152

Single-use mouthpiece (2 included)

PS-2522 (10-pack)

CD-ROM containing DataStudio workbook files

013-09084

Additional Equipment Required

PASPORT interface

See PASCO catalog or www.pasco.com

DataStudio or DataStudio Lite

(required for computer-based data collection)

See PASCO catalog or www.pasco.com
DS Lite available for free download

Additional Equipment Recommended

Additional single-use mouthpieces

PS-2522 (10-pack)

SP

IRO

M

ET

ER

ä

PS-2152

WAIT

READ

Y

Summary of Contents for PASPORT PS-2152

Page 1: ...ory volume and other important lung volumes and capacities Included Equipment Part Number Electronics enclosure and handle PS 2152 Single use mouthpiece 2 included PS 2522 10 pack CD ROM containing Da...

Page 2: ...collect and display data for a pulmonary function test On screen instructions will guide students through data collection and analysis Students can review the test procedure and interpretation of PFT...

Page 3: ...green READY indicator has illuminated place the tapered end of the mouthpiece between the test subjects s front teeth with the lips creating a seal to ensure that all inspired and expired air flows th...

Page 4: ...T indicator is blinking the sensor is measuring the ambient pressure in order to establish a baseline reading It is important not to expose the sensor to any pressure change or airflow during the wait...

Page 5: ...level of tidal expiration Inspiratory Reserve Volume IRV is the amount of air moved into the lungs during a forced inspiration beyond the level of tidal inspiration Forced Expiratory Volume in One Sec...

Page 6: ...forced expiration A typical RV is 1 2 L Functional Residual Capacity FRC is the amount of air left in the lungs after a normal expiration FRC RV ERV A typical FRC is 2 5 liters The formula for Lung V...

Page 7: ...apacity measurements taken from the graph since each measurement is made between two data points recorded within a few seconds In DataStudio you can eliminate the trend with this calculation Total Flo...

Page 8: ...tance with any PASCO product contact PASCO at Address PASCO scientific 10101 Foothills Blvd Roseville CA 95747 7100 Phone 916 786 3800 800 772 8700 Fax 916 786 3292 Web www pasco com Email techsupp pa...