background image

®

Model No. CI-6537

Force Sensor

5

Equipment Setup

Mounting the Force Sensor on a PASCO Dynamics Cart

The Force Sensor has two built-in mounting screws that align with the 
threaded holes in the accessory tray of a PASCO Dynamics Cart (such 
as the ME-9430 Plunger Cart or ME-9454 Collision Cart). The screws 
are spring loaded so they remain in a retracted position when not in 
use.

1. 

Position the sensor lengthwise it in the accessory tray of the Dynamics 
Cart.

2. 

Insert a size #0 Phillips head screwdriver into the threaded hole in 

the accessory tray of the Force Sensor, and align the screwdriver 
with the Phillips head screw. 

3. 

Press down with the screwdriver until the screw extends into the 

threaded hole on the dynamics car or cart.

4. 

Turn the screwdriver clockwise until the screw is tight. Repeat the 

process with the other screw. 

To mount other accessories (e.g. ME-9481 Bernoulli Cart Accessory) 
on top of the Force Sensor, attach the accessory in the Force Sensor 
accessory tray in the same way you would attach the accessory to a 
Dynamics Cart.

Mounting the Force Sensor on a Support Rod

The Force Sensor has a hole and thumbscrew at one end that allows 
you to mount the sensor on a support rod from 3/8” to 1/2” diameter. 

T

A R E

Figure 2:  Mounting the sensor on the Dynamics Cart

Summary of Contents for CI-6537

Page 1: ...Instruction Manual Manual No 012 05804C Force Sensor Model No CI 6537 CAUTION 50N 50N FORCE SENSOR...

Page 2: ...PASCO Interfaces 7 Suggested Experiments 8 Component of Force on an Incline Plane 8 Newton s Second Law Pushing and Pulling a Cart 8 Newton s Second Law Constant Force 9 Work Energy Theorem 9 Tension...

Page 3: ...798 3 Ziplock bag 1 NA Additional Equipment Recommended A PASCO ScienceWorkshop 500 or 750 interface CI 6400 or CI 6450 or CI 7599 A computer NA Introductory Dynamics System cart track track accessori...

Page 4: ...interface and a detachable hook The housing has a Tare button for zeroing the sensor on the same side of the housing as the cable and a thumbscrew for mounting on a support rod up to 1 2 diameter on t...

Page 5: ...ry tray of the Force Sensor and align the screwdriver with the Phillips head screw 3 Press down with the screwdriver until the screw extends into the threaded hole on the dynamics car or cart 4 Turn t...

Page 6: ...ce during an experiment set up the experimental equipment as needed and tare the sensor at the beginning of the experiment before taking data The sensor can maintain its zeroed condition for over thir...

Page 7: ...ctions in Appendix C Using the Force Sensor with PASCO Interfaces The instructions in this manual are intended for those using the Force Sensor with the PASCO 500 or 750 ScienceWorkshop interfaces Usi...

Page 8: ...eoretical value mgsin Newton s Second Law Pushing and Pulling a Cart When an object is accelerated by a net force the acceleration is directly proportional to the net force and inversely proportional...

Page 9: ...kinetic energy of the cart as it is pulled by a constant force Arrange the CI 6742 Motion Sensor CI 6537 Force Sensor and cart on the track as in the previous suggested experiment Set up a pulley stri...

Page 10: ...Accessory to the Dynamics Cart Arrange the CI 6742 Motion Sensor CI 6537 Force Sensor and friction cart on the track as in the previous suggested experiment Set up a pulley string and hanging mass so...

Page 11: ...or with the second force sensor Newton s Third Law Impulse Collision The impulse during a collision equals the change in momentum during the collision F t mv Mount the Force Sensor at one end of the t...

Page 12: ...o the speed length and mass of the pendulum Measure the change in mass of liquid nitrogen as it vaporizes versus the energy input to vaporize the liquid nitrogen Measure fluid drag forces on objects o...

Page 13: ...ms Bandwidth limit 2 kilohertz internal low pass filter Output drive 8 meters of cable without instability The range of the sensor is 50 newtons with an output between 8 to 8 volts or 160 millivolts p...

Page 14: ...In the Sensors list of the Experiment Setup window double click the Force Sensor icon to associate it with the same analog channel on the picture of the interface 4 Mount the Force Sensor vertically...

Page 15: ...mediately 5 Reapply the force Note If you have DataStudio open you can verify the zero measurement by taking data with no weight attached If the zero is not achieved press the Tare button again and en...

Page 16: ...you can hang a known mass from the hook Don t put any mass on the hook for this step d Double click on the Force Sensor icon to open the Sensor Setup dialog box The dialog box shows the default settin...

Page 17: ...Support For assistance with the CI 6537 Force Sensor or any other PASCO products contact PASCO as follows Address PASCO scientific 10101 Foothills Blvd Roseville CA 95747 7100 Phone 916 786 3800 FAX 9...

Page 18: ...a period of one year from the date of shipment to the customer PASCO will repair or replace at its option any part of the product which is deemed to be defective in material or workmanship The warran...

Reviews: