Page 2-6
MiniCapt Remote Microbial Air Sampler Operations Manual
Chapter 2 Unpacking and Installation
Installation
Installing the Electronics Enclosure
The MiniCapt Remote Microbial Air Sampler:
• Is designed to be mounted to a wall or other vertical surface that can support 10 lb
(4.5 kg).
• Must be installed by qualified personnel.
The enclosure may be mounted on laboratory structure material such as Unistrut
®
using similar attachment methods as for mounting on a metal framed hollow wall or
solid metal sheet wall.
Attaching the Chassis to the Wall
You will need four metal fasteners with a diameter of 5 – 6 mm. See
on page 2-7 for more details.
>>
To prepare a wood or metal stud-wall for chassis installation:
1
Using the mounting slots/hole on the back of the chassis as a template, mark the
drill points.
2
Drill pilot holes of a size appropriate to the size fastener you are using.
>>
To prepare a sheet metal wall for chassis installation:
The sheet metal wall must be a minimum of 3 mm thick aluminum or steel.
1
Using the mounting slots/hole on the back of the chassis as a template, mark the
drill points.
2
Drill and tap the wall for the size of fastener you will be using.
The screws must be long enough to pass completely through the wall material.