Page 3-4
HandiLaz Mini Operator’s Manual
Chapter 3: MEASURE MODE Settings
Taking SINGLE Sample
1
From the SAMPLING display, press the
START
/
STOP
key.
The initial delay countdown will begin.
2
After the initial delay, press the
START
/
STOP
key.
Using CONT (Continuous) Sampling Mode
Use this sampling mode to continuously sample until the operator stops the
HandiLaz Mini by pressing the
START
/
STOP
key. The maximum duration of a
continuous sample is 99 minutes and 59 seconds.
To open the CONT sampling mode:
1
From the main menu, use either arrow (
▲
or
▼
) key to navigate to MEASURE
MODE.
2
Press the
ENTER
key.
The MEASURE MODE menu will open.
3
Use either arrow key to navigate to CONT.
4
Press the
ENTER
key.
The CONT measure mode setup screen will open.
Setting CONT (Continuous) Sample Parameters
The parameters seen when the CONT measure mode is opened are those that
were saved when the CONT mode was last run. The CONT measuring mode
can be run as it is, without changing any sample parameters.
To edit the setup screen:
1
Press the
ENTER
key.
Pressing the
ENTER
key moves the cursor in a loop through each field that can
be edited.
2
At the field to be changed, press the up
▲
or down
▼
key.
The field setting will loop through its range of allowable settings.
3
When the field is set correctly, press the
ENTER
key again to move to the next
field that can be edited.
SINGLE 0.3/0.5/5.0
µ
m
LOC.023 STR:N BEEP:N
FREE REC.:09950
NO