Maintenance
Clean the CyFlow
®
Cube casing on a regular base carefully with soft cloth. Water must not enter the
CyFlow
®
Cube or peripheral devices or come into contact with electric connections and switches. For
cleaning the screen, always use special screen cleaner and soft cloth.
Do not use any organic solvents, nitro thinner, benzol, alcohol, highly concentrated bleach etc!
For cleaning of flow cuvette, refer to the described cleaning procedure. Do not use tools to clean the flow
cuvette. In case the flow cuvette is blocked, enquire Partec for rapid exchange.
Regularly empty the waste bottle and clean with warm detergent solution and a brush.
Clean sheath reservoir with distilled water and a clean brush and flush with clean distilled water several
times.
Remember: a clean sheath fluid reservoir is critical for proper operation.
If the CyFlow
®
Cube will not be used for longer periods, clean flow system by using distilled water. Put a
sample tube half-ways filled with distilled water at the sample port. Clean waste and sheath reservoir, wipe
top dry.
Service
All service is to be made from an authorized service engineer. Please contact your local supplier or Partec
([email protected]).
For further information, please consult our web site:
Products
Application tips
Instrument and software support
News and events
And much more ...
www.partec.com
Warning: The Waste may contain biohazardous and carcinogenic
material from the samples (infectious material, dyes).