![Parr Instrument Company 6300 Operating Instructions Manual Download Page 23](http://html1.mh-extra.com/html/parr-instrument-company/6300/6300_operating-instructions-manual_3774715023.webp)
P
ROGRAM
I
NSTALLATION
& C
ONTROL
6300
4
w w w . p a r r i n s t . c o m
23
C
HAPTER
4
P
ROGRAM
I
NSTALLATION
&
C
ONTROL
S
OFTWARE
I
NSTALLATION
The program in the 6300 Calorimeter can be extensively
modified to tailor the unit to a wide variety of operating
conditions, reporting units, laboratory techniques,
available accessories and communication modes.
In addition, the calculations, thermochemical
corrections and reporting modes can be modified to
conform to a number of standard test methods and
procedures.
Numerous provisions are included to permit the use of
other reagent concentrations, techniques, combustion
aids and short cuts appropriate for the user’s work.
Note:
Changes to the program are made by use of the menu
structure described in Appendix A of this manual. Any
of these items can be individually entered at any time to
revise the operating program.
D
EFAULT
S
ETTINGS
Units are pre programmed with DEFAULT
SETTINGS. See pages 24 and 25 for a listing of the
factory default settings.
These default settings remain in effect until changed
by the user. Should the user ever wish to return to the
factory default settings, go to the Program Information
and Control Menu, then to User/Factory Settings, and
then touch Reload Factory Default Settings and YES.
Non-volatile memory is provided to retain any and
all operator initiated program changes; even if power
is interrupted or the unit is turned off. If the unit
experiences an intentional or unintentional “Cold
Restart”, the controller will return to its default settings.
R
EVISING
D
EFAULT
S
ETTINGS
The default parameters of the 6300 Calorimeter can
be changed to guarantee that the 6300 Calorimeter,
when cold restarted, will always be in the desired
configuration before beginning a series of tests.
Users who wish to permanently revise their default
settings may do so using the following procedure:
Establish the operating parameters to be stored as
•
the user default settings.
Go to the Program Info and Control Menu, User/
•
Factory Settings, User Setup ID, and enter the
desired User Setup ID.
Select Save User Default Settings
•
To re-load the user default settings, go to the Program
Info and Control Page, User/Factory Settings, Re-load
User Default Settings, and YES.
Summary of Contents for 6300
Page 2: ......
Page 10: ...PREFACE P a r r I n s t r u m e n t C o m p a n y 10 This page left blank intentionally...
Page 20: ...2 INSTALLATION P a r r I n s t r u m e n t C o m p a n y 20 This page intentionally left blank...
Page 76: ...B CALCULATIONS P a r r I n s t r u m e n t C o m p a n y 76 This page intentionally left blank...
Page 123: ...PARTS LISTS DRAWINGS 6300 F w w w p a r r i n s t c o m 123 Figure F 20 6300 Air Can Assembly...
Page 129: ......