9
Specifications - Sensors
Power
5 VDC provided by SciTemp
®
Monitor.
Fluid Connections
Choice of five sizes: Luer, 3/8” Barb, 1/2”
Barb, 3/4” Tri-Clover (TC) Sanitary and 1” Tri-
Clover (‘Ladish’) Sanitary.
Wetted Materials
Polysulphone meets USP Class VI and FDA
21 CFR 177.1520, all wetted materials made
with animal-free compounds.
Sensor Type
Thermistor, Epoxy coated, 2252
Ω
.
Temperature Range
-40 to +150° Celsius.
Accuracy
+/- 0.10° Celsius in the 4.0 to 70° Celsius
range.
Resolution
0.01 ° Celsius
Sensor Microchip
EPROM, Stores Sensor ID, and
Calibration Factor.
Sensor Connector & Cables
Lockable and Waterproof.
Summary of Contents for SciLog SciTemp