Parker SciLog SciTemp Installation, Operation & Maintenance Instructions Manual Download Page 6

 

 

Introduction: 

You will find the SciTemp® Temperature Sensor System easy to use. The state-of-the-art 
hardware and software design of the SciTemp® Temperature Sensor and Monitor combination 
allows you to measure and document the Temperature of many filtration / separation processes. 
With proper maintenance, the SciTemp® Monitor will provide many years of excellent service 
and performance

Please read the following instructions carefully! 

Inspections:

 Remove the products carefully from the shipping container. Check the contents 

against the purchase order to verify that all parts are included and undamaged. 

Please do the inspection now, even if the products are not used immediately. Many carriers 
must receive damage claims within seven days of delivery. Please retain all packing material so 
unit may be shipped safely, if necessary. 

Customer Service: 

Parker customer service:  If assistance is required, please contact us at: 

Parker Hannifin Manufacturing Ltd  

Parker Hannifin Corporation

 

Bioscience Division – EMEA 

Bioscience Division – N.A

 

Durham Road 

2340 Eastman Avenue

 

Birtley, Co. Durham 

Oxnard, California, USA 93030

 

DH3 2SF, England 

toll free: 877 784 2234

 

phone +44 (0)191 4105121 

phone: +1 805 604 3400

 

fax +44 (0)191 4105312 

fax: +1 805 604 3401

 

email:

 

[email protected] email: [email protected] 

www.parker.com/bioprocessing www.parker.com/bioprocessing 

Parker customer service personnel will be able to serve you more efficiently if you have the following 
information: 

  Serial number and model name of the equipment 

  Installation procedure being used 

  Concise list of symptoms 

  List of operating procedures and conditions in use when problem arose 

Calibration: 

The SciTemp® Temperature Monitor is a calibrated device, and is calibrated with test 
equipment that is traceable to NIST.   
A Calibration Certificate is included with each unit, and has a one year expiration date.   
It is highly recommended that the monitor be calibrated on an annual basis or more often if 
your Metrology Department deems it necessary. 
Parker provides this Calibration Service, complete with a new certificate, showing the “as 
found” and “as left” data.   
Contact local Parker representative. 

Summary of Contents for SciLog SciTemp

Page 1: ...1 SciLog SciTemp Temperature Monitor Sensor Installation Operation Maintenance Instructions...

Page 2: ...ormed of any alterations This publication is for general information only and customers are requested to contact our Process Filtration Sales Department for detailed information and advice on a produc...

Page 3: ...11 2 0 Front Panel Data Entry Display 12 3 0 Back Panel Interface Options 13 4 0 Disposable Temperature Sensors 16 Part B SciTemp Temperature Sensor Monitor Software 17 1 0 Software Overview 17 2 0 Ma...

Page 4: ...ING SERVICING CLEANING OR REMOVING THE FUSE FAILURE TO DO SO COULD RESULT IN BODILY HARM AND OR PROPERTY DAMAGE CAUTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES WARNING ONLY PE...

Page 5: ...RETIRER LE FUSIBLE NE PAS OBSERVER CES PR CAUTIONS RISQUERAIT DE CAUSER DES BLESSURES CORPORELLES OU ET D ENDOMMAGER L QUIPEMENT PRUDENCE SOYEZ PRUDENT LORSQUE VOUS MANIPULEZ DES APPAREILS SENSIBLES...

Page 6: ...tion Bioscience Division EMEA Bioscience Division N A Durham Road 2340 Eastman Avenue Birtley Co Durham Oxnard California USA 93030 DH3 2SF England toll free 877 784 2234 phone 44 0 191 4105121 phone...

Page 7: ...ved See SciTemp Temperature Sensor Specifications The protection provided by this equipment may be impaired if the SciTemp Temperature Sensor is used in a manner inconsistent with this manual or for p...

Page 8: ...e damaged if mishandled I O Port 2 Output Female DB25 four TTL outputs for alarms four 4 20 ma Analog output signals 18 bit resolution I O Port 3 Printer PC DB9FM RS 232 output I O Port 3 Serial I O F...

Page 9: ...tted Materials Polysulphone meets USP Class VI and FDA 21 CFR 177 1520 all wetted materials made with animal free compounds Sensor Type Thermistor Epoxy coated 2252 Temperature Range 40 to 150 Celsius...

Page 10: ...custom macros for data compilation Includes Winwedge interface software Sent to you ready to use Complete process analysis with graphing of data Real time verification and documentation of process pa...

Page 11: ...the differential temperature The monitor s output signals are in both analog and digital formats It comes with a small 110 VAC wall adapter as a power supply The disposable SciTemp Temperature Sensor...

Page 12: ...and programming of the SciTemp Temperature Monitor The basic key definitions are MENU Allows entry into the parameter settings menu ENTER Accepts the selected parameter setting EXIT Exits the menu re...

Page 13: ...ature sensors are connected to these two using the included cables 3 2 Output Port Provides two 4 20 mA analog outputs 18 bit resolution and TTL outputs for the 4 alarms Hi Low for each temperature ch...

Page 14: ...eeded in the loop connect pin 1 Vpwr to pin 14 4 20 and connect the output cable to pin 2 4 20 and pin 15 Gnd For 4 20 ma output of Temperature at T2 Analog 2 use pins 4 4 20 and 16 4 20 If excitation...

Page 15: ...ow Alarm TTL 2 Pins 24 Out1 and 13 Gnd P3 Hi Low Alarm TTL 3 Pins 12 Out2 and 13 Gnd TM or DP Hi Low Alarm TTL 4 Pins 25 Out3 and 13 Gnd TTL outputs are either 0 or 5 VDC and carry minimal current The...

Page 16: ...7 1520 All Wetted Materials are made of Animal Free Compounds Compatible with Most Sanitizing Agents such as NaOH Hypochlorite Sensors can be repeatedly autoclaved Sensor Type Thermistor Epoxy Coated...

Page 17: ...tial Temperature The Main Menu allows you to change the temperature units C or F the high and low analog range settings adjust the baud rate and print time as well as configure user defined Hi Low tem...

Page 18: ...18...

Page 19: ...to set the data output interval 1 to 60 sec Default 5 sec 2 7 HI ALARM T1 Used to set High Temperature Alarm value should be within High and Low Range set in 2 3 and 2 4 above Hi Alarms exist for bot...

Page 20: ...20...

Page 21: ...press Exit 3 3 CUSTOM CAL SENSOR When you reach this screen and are ready to perform the custom calibration at the upper limit Be sure you have verified the temperature of the calibration solution us...

Page 22: ...8 21 72 21 96 10 53 23 21 72 21 96 10 53 28 21 72 21 96 10 53 33 21 72 21 96 10 53 38 21 72 21 96 10 53 43 21 72 21 96 10 53 48 21 72 21 96 10 53 53 21 72 21 96 10 53 58 21 72 21 96 10 54 03 21 72 21...

Page 23: ...received It also automatically populates a graph in real time as the data is generated Discussion of this package begins on the next page Alternatively when a PC is connected to this serial port all d...

Page 24: ...or to an available Com Port on your PC Minimum system requirements are Windows 98 Excel 2000 and WinWedge32 v3 0 which is included in the package Once installed click on the shortcut for the spreadshe...

Page 25: ...with the next screen Enter all of the appropriate Operator Information in this dialog box and click on OK PLEASE NOTE the Clear Data button does not affect the information in these fields You may cli...

Page 26: ...data The text box in the upper left of the spreadsheet contains the header information that is generated by the SciTemp Temperature Sensor and Monitor This will list the version the calibration data a...

Page 27: ...ta worksheet cells are protected the charts are not with this purpose in mind When you are finished with a run your screen will look similar to this one Click on File Save as and choose an appropriate...

Page 28: ...ature Sensor Press Ok 2 From the Phone Number screen select Direct to Com 1 in the box labeled Connect Using or select any other available Com port Press Ok 3 From the Com 1 Property screen select the...

Page 29: ...water 2 SciTemp sensors can be repeatedly sanitized with any of the listed sanitizing solutions Do not exceed the recommended number of sanitization cycles Sensor accuracy and precision will be affect...

Page 30: ...30...

Page 31: ...31...

Page 32: ...o Durham DH3 2SF England phone 44 0 191 4105121 fax 44 0 191 4105312 email bioscience emea parker com www parker com bioprocessing Parker Hannifin Corporation Bioscience Division N A 2340 Eastman Aven...

Reviews: