24
Command List:
Each command must be followed by a carriage return and line feed character:
A
ABORT:
this stops the dispensing cycle immediately.
Vxxx
VOLUME
to be dispensed. This is a three digit value. For example, send V050 for 5.0 ml (50
mg), V100 for 10.0 ml (100 mg). The dispensing cycle runs in the current pump direction.
Wxxxx
WEIGHT
to be dispensed. This requires a connection between the pump and a balance.
This is a four digit value. For example to dispense 2.00 grams, send W0200.
X
REPEAT:
begins another dispensing cycle for the current dispense volume or weight.
F
FORWARD:
sets the current pump direction to clockwise, CW will be displayed.
R
REVERSE:
sets the current pump direction to counter-clockwise, CCW will be displayed.
P
PRIME:
primes the system by running forward for a programmed time interval “T”.
E
EMPTY:
empties the system by reversing the flow back to the bottle for a programmed time
interval
“T”.
Cxx
COMMAND:
sets the Prime/Empty time to a value between 1-99. This must be a two digit
number.
T
TIME:
current time interval for the Prime/Empty cycle.
S
STATUS:
the system reports its status.