background image

 

Cylinder Operation / Maintenance 

Cylinder  Pressure  Readings:

  Oxygen  is  a  true 

compressed  gas,  while  in  the  cylinder,  thus  the 
cylinder pressure gauge can be used to determine the 
amount of gas remaining in the cylinder. For example, 
2000 psi indicates full, 1000 psi indicates half full, etc. 
Nitrous  Oxide  is  a  liquefied  compressed  gas  that 
vaporizes  in  the  cylinder,  thus  the  cylinder  pressure 
gauge cannot be used to determine the amount of gas 
remaining in the cylinder until all liquid in the cylinder 
vaporizes.  While  liquid  remains  in  the  cylinder,  the 
cylinder pressure gauge indicates the vapor pressure 
which depends on and varies with the temperature of 
the liquid. For example, at 68ºF, the vapor pressure is 
about 750 psi; at 20ºF, it drops to about 400 psi; while 
at  90ºF,  it  increases  to  about  1000  psi.  After  all  the 
liquid  vaporizes,  the  pressure  will  decrease  normally 
as  the  gas  is  withdrawn,  and  the  cylinder  pressure 
gauge can then be  used to determine the  amount of 
gas remaining in the cylinder. 

Cautions 

 

Always turn on CYLINDER VALVES slowly and 
fully. 

 

NEVER  ATTEMPT TO LOOSEN  cylinder  valve 
packing  nut.  If  valve  stem  is  tight,  return 
cylinder to supplier. 

 

Warning:  Do  not  remove  or  alter  gas 

indexing pins 

Good Practices: Cylinders with Mobile Cart

 

1.  Two  cylinders  of  N

2

O,  or  one  cylinder  of  O

2

  and 

one cylinder of N

2

O are typically connected.   

 

2.  Minimize leak risks: Confirm Yoke Washers are in 

place  before  replacing/mounting  cylinders.  Use 
Porter  #PA-1656-000  replacement  washers 
(once/yr. or as needed). Have spare washers. 

 

3.  Minimize leak risks: Tighten Tee Handle securely, 

assuring washer is in flat sealing position and yoke 
pins are in place in cylinder post holes.  

 

4.  Assure Mobile Cart is populated with cylinders of 

N

2

O,  or  O

2

  and  N

2

O  with  adequate  degree  of  fill 

before starting any procedure. 

 

5.  Label cylinders 

with a tag or sticker indicating “In-

Use” and “Full” (“Full” is reserve), if appropriate. 

 

6.  Use  Cylinder  W

rench  to  open  the  “In-Use” 

cylinders.  Verify  wrench  is  attached  to  Cart  (at 
Utility Basket).   

 

7.  Cylinder 

pressure 

gauges 

on 

Regulator 

Assemblies provide a visual indication of cylinder 
status  (see  details  on  Cylinder  Pressure 
Readings) 

 

8. 

When  “In-Use”  cylinder  is  depleted,  close  valve, 
bleed  gas  by  slowly  loosening  Tee  Handle,  and 
replace  empty  cylinder.    Connect  the  Regulator 
Assembly to 

“Full” cylinder and open valve.   

 

9. 

When “In-Use” O

2

 cylinder is depleted, the Oxygen 

Fail Safe will stop N

2

O flow through Flowmeter or 

Nitronox automatically. 

 

10. 

When  “In-Use”  N

2

O  cylinder  is  depleted,  the 

Flowmeter or Nitronox will deliver 100% O

2.

  

 

11.  After each use, turn off cylinder valves.   

 

12.  Method  to  use  trapped  nitrous  oxide: Turning  off 

cylinder valves and turning Flowmeter or Nitronox 
flow valves to off positions (or turning on/off valve 
to off) will trap a small amount of nitrous gas in the 
gas  supply  hose  (assisted  by  check  valve  in 
nitrous gas Regulator Assembly). Starting the next 
procedure  by  turning  on  cylinder  valves  and 
starting  flow  through  the  Flowmeter  will  use  this 
trapped gas (as opposed to venting the gas to the 
environment).  

 
Monthly Check 
Flowmeter  Leak  Test.

  Attach  the  gas  supply  Hoses 

to a Flowmeter and turn the flow control knobs to the 
OFF  position  and  the  ON  /  OFF  switch  to  the  OFF 
position. 
Turn  cylinders  ON.  Pressurize  the  gas  supply  lines 
and  Mobile  Cart  Regulator  Assemblies  with  cylinder 
pressure. Turn cylinders OFF. Apply masking tape to 
both  gauge  faces.  Tap  lightly  on  gauges  and  mark 
gauge  needle  positions  on  the  masking  tape.  There 
should be little or no movement of the gauge needles 
after 15 minutes. 

Nitronox Leak Test:

 Leave O2 supply on, close N2O 

cylinder valve off, and apply tape only on N2O gauge 
as  above  and  watch  for  movement  of  gauge  needle 
after 15 minutes.  

 

 

 

Summary of Contents for Porter Nitronox

Page 1: ...ry Technique Safety Checks Maintenance and Troubleshooting Proper Use of Cylinders Connection to Flow Devices Use FM 1345 Assembly Instructions Examination of Contents Examine shipping carton for sign...

Page 2: ...e 1 5 Basket 1 6 Basket Clips 2 7 Cylinder Holder 1 8 Set Screw 1 9 Cylinder Wrench w Chain 1 Item Component Description Qty Image 10 Mounting Adapter 1 11 End Cap w 3 Screws 1 12 Hex Nut Adapter 1 13...

Page 3: ...3...

Page 4: ...rom the suspected crossed lines area WARNING Porter Instrument equipment utilizes the cross protection system The flexible hose and connectors that connect to the Regulator Assembly and the Flowmeter...

Page 5: ...rs A Loaded Mobile Cart must return to vertical when tipped 10 Do not roll cart aggressively For hardware protection follow Regulator Assembly configuration and location of nitrous oxide cylinder with...

Page 6: ...4 2 Check the color of the regulator gauges to ensure the O2 gauge is green white for ISO model the N2O gauge is blue 3 Check the indexing pins on each regulator to ensure they re indexed for the cor...

Page 7: ...m 12 screwed into the device mounting feature See Fig 7 illustration for mounting feature of Nitronox 2 Tighten the Hex Nut Adapter to lock the Mounting Adapter into place 3 Place the Mounting Adapter...

Page 8: ...Assembly see Fig 11 5 Slide the Regulator Assembly into place by lining up the pins and pin holes on the cylinder Push into place Properly placed the Regulator Assembly should be supported on the pin...

Page 9: ...adequate degree of fill before starting any procedure 5 Label cylinders with a tag or sticker indicating In Use and Full Full is reserve if appropriate 6 Use Cylinder Wrench to open the In Use cylind...

Page 10: ...he Utility Basket Note If location dimensions of the Utility Basket and Cylinder Holder were not strictly followed see FM 1345 2 Cylinder Mobile Cart Assembly Instructions the Scavenger Interface may...

Page 11: ...1 damaged or misused 2 repaired altered or modified by persons other than Parker Hannifin Corporation 3 not installed in strict compliance with applicable codes and ordinances or 4 not installed by a...

Reviews: