WARNING:

  External  gas  supplies  may  range  from  40  to  65  psi  line  pressure.

   

Follow  safe  connection 

protocol  in  this  order: 

Connect  external  gas  supply  hoses  to  DISS  (Diameter  Indexed 

Safety  System)  fittings  of  Flowmeter  or  Nitronox  (O2  first); 

then  connect

  to  external 

supply  source.  Do  not  disconnect  with  supply line  pressurized.  Disconnect  at  supply 
source first, per appropriate safe protocol. 

 

WARNING: 

During any power outage, remember to turn OFF the flowmeter and manually turn OFF the 

cylinder valves (or external supplies).  With centralized, electrically  powered gas systems, if 
gas  was  flowing  when  the  power  went  out  and  the  flowmeter  is  left  ON,  gas  will  be  flowing 
when the power is restored. 

 

CAUTION:

 Always  use clean,  dry medical grade gases.  Introduction of moisture or  other contaminants 

into the system and flowmeter device may result in defective operation. 

 

CAUTION: 

Do not attempt to repair, alter or calibrate the system or device (except as directed in Porter 

User’s Manuals).  Unauthorized repair, alteration or misuse of the Flowmeter or Nitronox device 
is likely to adversely affect the performance and will void the warranty. 

 

CAUTION:

 

Always  lock  casters  in  place  when  2  Cylinder  Mobile  Cart  is  not  being  transported  to 

prevent  cart  from  moving  during  use  or  maintenance.    Protect  against  tip  over  and 
potential damage to high pressure gas cylinders. A Loaded Mobile Cart must return to 
vertical when tipped 10°. Do not roll cart aggressively. For hardware protection, follow 
Regulator Assembly configuration and location of nitrous oxide cylinder within Cylinder 
Holder as described in Proper Attachment of Gas Cylinders. 

 
CAUTION:

 

Never oil or grease any part of this system (minimize fire or explosion potential). 

 
 

INTENDED USE: 
Flowmeter or Nitronox for a nitrous oxide & oxygen delivery system: 

The device is intended for patient use and delivers a mixture of nitrous oxide and oxygen gases 
for a procedure in a medical healthcare setting by trained medical personnel.  Porter Instrument 
recommends the user be thoroughly familiar with the use of Nitrous Oxide - Oxygen delivery for 
patient analgesia and be properly trained in its administration prior to using this product.   
 

INTENDED USE: (Accessory) 
A nitrous oxide & oxygen delivery system with Mobile Cart: 

 
The Porter 2-

Cylinder Mobile Cart is for use with one “E” cylinder of O2 and one “E” cylinder of 

N2O, or two cylinders of N2O when using other O2 gas supply.  The two pressure gauges reflect 
cylinder  pressure  only.    When  the  gas  supply  is  opened,  gas  will  flow  through  the  portable 
regulatory assembly and be regulated down to 53 PSIG at the output of the regulators, out through 
the hoses to the Flowmeter or Nitronox device. A check valve within the N2O regulator assembly 
assists to minimize trapped gas venting out into the room after procedures are completed.  Each 
Mobile Cart has a fixed 44" height (w/o Flowmeter or Nitronox attached) for use or storage.   

 
 

Summary of Contents for Porter Nitronox

Page 1: ...ry Technique Safety Checks Maintenance and Troubleshooting Proper Use of Cylinders Connection to Flow Devices Use FM 1345 Assembly Instructions Examination of Contents Examine shipping carton for sign...

Page 2: ...e 1 5 Basket 1 6 Basket Clips 2 7 Cylinder Holder 1 8 Set Screw 1 9 Cylinder Wrench w Chain 1 Item Component Description Qty Image 10 Mounting Adapter 1 11 End Cap w 3 Screws 1 12 Hex Nut Adapter 1 13...

Page 3: ...3...

Page 4: ...rom the suspected crossed lines area WARNING Porter Instrument equipment utilizes the cross protection system The flexible hose and connectors that connect to the Regulator Assembly and the Flowmeter...

Page 5: ...rs A Loaded Mobile Cart must return to vertical when tipped 10 Do not roll cart aggressively For hardware protection follow Regulator Assembly configuration and location of nitrous oxide cylinder with...

Page 6: ...4 2 Check the color of the regulator gauges to ensure the O2 gauge is green white for ISO model the N2O gauge is blue 3 Check the indexing pins on each regulator to ensure they re indexed for the cor...

Page 7: ...m 12 screwed into the device mounting feature See Fig 7 illustration for mounting feature of Nitronox 2 Tighten the Hex Nut Adapter to lock the Mounting Adapter into place 3 Place the Mounting Adapter...

Page 8: ...Assembly see Fig 11 5 Slide the Regulator Assembly into place by lining up the pins and pin holes on the cylinder Push into place Properly placed the Regulator Assembly should be supported on the pin...

Page 9: ...adequate degree of fill before starting any procedure 5 Label cylinders with a tag or sticker indicating In Use and Full Full is reserve if appropriate 6 Use Cylinder Wrench to open the In Use cylind...

Page 10: ...he Utility Basket Note If location dimensions of the Utility Basket and Cylinder Holder were not strictly followed see FM 1345 2 Cylinder Mobile Cart Assembly Instructions the Scavenger Interface may...

Page 11: ...1 damaged or misused 2 repaired altered or modified by persons other than Parker Hannifin Corporation 3 not installed in strict compliance with applicable codes and ordinances or 4 not installed by a...

Reviews: