18
START C
1
4-20mA output
Disconnect receiving
instrument. Connect an
ampere meter directly
to the output terminals.
Can you read current
output?
Either:
-Wiring broken
-Malfunctioning of
receiving instrument
-Non conformity of input
and output
Output circuit
failure
Contact Technical
Support Specialist
Perform loop check
Is flow direction
correct?
Correct flow
direction
Correct combination of detector
and converter? Is detector size
set correctly at converter?
Make pairing
correct
Is detector filled
with liquid?
Relocate detector
to make it filled
with liquid
Is zero point
adjustment correct?
Stop flow and
adjust zero point
YES
NO
NO
NO
NO
NO
YES
YES
YES
Troubleshooting
C:
When no signal is present while running the fluid.