15
6.3 The controller display will not illuminate
6.4 Excessive downstream temperature
Indication
Possible fault / Condition
Action
-
The generator is not powered
Check the following:
1
the generator is connected to a 'Live' supply.
2
the ON / OFF switch is in the 'I' position.
3
the fuse in the IEC320 socket and the fuse in
the plug of the supply cable (UK only).
4
the electrical supply cable is not damaged.
Contact Parker Hannifin.
Faulty controller or internal wiring fault.
Contact Parker Hannifin.
Indication
Possible fault / Condition
Action
Downstream monitoring.
Faulty cooling fan.
Contact Parker Hannifin.
The vents on the chassis of the generator are
blocked
Check that the vents are not blocked and air is
flowing correctly.
Inlet / outlet ports incorrectly connected
Check that the generator is correctly connected.
The flow of air through the generator is too low.
Verify that the flow rate matches that specified in
section 2.1 for the generator.