2 Description
Parker Hannifin zero air generators are designed to produce a supply of hydrocarbon free air from a clean, dry (IS0 8573-1:2010 class 2.2.1)
compressed air source so eliminating the need for high pressure gas storage and disruption due to bottle changes and subsequent re-
calibrations.
The inlet compressed air passes through a 1.0 micron filter to remove any particulate and then over a heated platinum catalsyst. This catalytic
action produces an ultra pure air stream with a residual hydrocarbon concentration (as methane) of less than 0.1 ppm. The gas exits the catalyst
unit, and is then cooled, via a cooling coil. Finally, the air passes through a 0.01 micron filter to remove any particulate and water, which may be
generated during the hydrocarbon removal phase.
2.1 Technical Specification
This specification is valid when the equipment is located, installed, operated, and maintained as specified within this user guide.
Parameter
Units
UHP
10ZA-S
35ZA-S
50ZA-S
75ZA-S
150ZA-S
200ZA-S
300ZA-S
General
Inlet Air Quality
ISO 8573-1: 2010
Class 2.2.1
Inlet Pressure
bar g
(psi g)
4 - 10
(58 - 145)
Inlet Temperature
o
C
(
o
F)
5 - 40
(41 - 104)
Flow Rate
L/min
1
3.5
5.0
7.5
15
20
30
Purity
ppm
< 0.1
(methane - with 100 ppm challenge)
Mechanical Connections (Compression fittings)
Compressed Air Inlet
1/8”
1/4”
Zero Air Outlet
1/8”
1/4”
Electrical Data
Connection Type
IEC 320
Input Voltage
Vac
104 - 127 60Hz
Power
W
55
160
575
Fuse
1
1.
Anti surge (T), 250V, 5 x 20mm HBC, Breaking Capacity 1500A @250V, IEC 60127, UL R/C
A
1
2
6.3
Input Voltage
Vac
207 - 253 50/60Hz
Power
W
55
230
575
Fuse
1
A
1
2
3
Environmental Data
Ambient Temperature
o
C
(
o
F)
5–40
(41–104)
Relative Humidity
50% @ 40
o
C (104
o
F) (80%
MAX
≤
31
o
C (87.8
o
F))
IP Rating
IP20, NEMA 1, indoor use only
Pollution Degree
2
Installation Over voltage
Category
II
Maximum Altitude
m
(ft.)
2000
(6562)
Noise
dB(A)
<50