2 Description

Parker Hannifin zero air generators are designed to produce a supply of hydrocarbon free air from a clean, dry (IS0 8573-1:2010 class 2.2.1) 
compressed air source so eliminating the need for high pressure gas storage and disruption due to bottle changes and subsequent re-
calibrations.
The inlet compressed air passes through a 1.0 micron filter to remove any particulate and then over a heated platinum catalsyst. This catalytic 
action produces an ultra pure air stream with a residual hydrocarbon concentration (as methane) of less than 0.1 ppm. The gas exits the catalyst 
unit, and is then cooled, via a cooling coil. Finally, the air passes through a 0.01 micron filter to remove any particulate and water, which may be 
generated during the hydrocarbon removal phase.

2.1 Technical Specification

This specification is valid when the equipment is located, installed, operated, and maintained as specified within this user guide.

  

Parameter

Units

UHP

10ZA-S

35ZA-S

50ZA-S

75ZA-S

150ZA-S

200ZA-S

300ZA-S

General 

Inlet Air Quality

ISO 8573-1: 2010

 Class 2.2.1

Inlet Pressure

bar g

(psi g)

4 - 10

(58 - 145)

Inlet Temperature

o

C

(

o

F)

5 - 40

(41 - 104)

Flow Rate

L/min

1

3.5

5.0

7.5

15

20

30

Purity

ppm

< 0.1

(methane - with 100 ppm challenge)

Mechanical Connections (Compression fittings)

Compressed Air Inlet 

1/8”

1/4”

Zero Air Outlet

1/8” 

1/4” 

Electrical Data

Connection Type

IEC 320

Input Voltage

Vac

104 - 127  60Hz

Power

W

55

160

575

Fuse

1

1.

Anti surge (T), 250V, 5 x 20mm HBC, Breaking Capacity 1500A @250V, IEC 60127, UL R/C

A

1

2

6.3

Input Voltage

Vac

207 - 253  50/60Hz

Power

W

55

230

575

Fuse

1

A

1

2

3

Environmental Data

Ambient Temperature

o

C

(

o

F)

5–40

(41–104)

Relative Humidity

50% @ 40

o

C (104

o

F) (80% 

MAX

 

 31

o

C (87.8

o

F))

IP Rating

IP20, NEMA 1, indoor use only

Pollution Degree

2

Installation Over voltage 
Category

II

Maximum Altitude

m

(ft.)

2000

(6562)

Noise

dB(A)

<50

Summary of Contents for domnick hunter UHP10ZA-S

Page 1: ...ro Air Generator UHP 10ZA S UHP 300ZA S Original Language User Guide aerospace climate control electromechanical filtration fluid gas handling hydraulics pneumatics process control sealing shielding E...

Page 2: ...ccordance with such instruction and with good practice The company shall not be under any liability whatsoever under the warranty if before giving notification in writing to the company as aforesaid t...

Page 3: ...layout 8 3 1 1 Installation Parts 8 3 2 Connecting the generator 9 3 2 1 Electrical supply 9 3 2 2 Stacking the Generator 9 4 Operating the equipment 10 4 1 Overview of the generator 10 4 2 Starting t...

Page 4: ......

Page 5: ...ed area and all of the vent ports on the rear of the generator are kept clear and free from blockages Use of the equipment in a manner not specified within this user guide may result in an unplanned r...

Page 6: ...be worn Gloves must be worn Warning Generator must be shut down and depressurised before performing any maintenance Refer to manual Caution Hot components enclosed Do not remove the cover until the u...

Page 7: ...cification is valid when the equipment is located installed operated and maintained as specified within this user guide Parameter Units UHP 10ZA S 35ZA S 50ZA S 75ZA S 150ZA S 200ZA S 300ZA S General...

Page 8: ...flicker in public low voltage supply systems for equipment with rated current 16 A per phase and not subject to conditional connection This equipment has been tested to and complies with the following...

Page 9: ...A S 325 12 8 340 13 4 425 16 7 219 8 6 10 2 22 5 35ZA S 455 17 9 340 13 4 425 16 7 330 13 0 14 2 31 3 50ZA S 455 17 9 340 13 4 425 16 7 330 13 0 14 2 31 3 75ZA S 455 17 9 340 13 4 425 16 7 330 13 0 14...

Page 10: ...s final location installation site and left to stabilise prior to unpacking Failure to do this could cause condensing humidity and potential failure of the generator 2 5 2 Unpacking Once ready to inst...

Page 11: ...llowed Class 1 Oil In each cubic metre of compressed air not more than 0 01 mg of oil is allowed Note This is the combined level for aerosol liquid and vapour ISO 8573 1 2010 Class 2 2 1 can be achiev...

Page 12: ...XPFA BP 1 8 OD Tube Non Return Valve 2A C2L 1 BN BP 1 4 OD Tube Non Return Valve 4A C4L 1 BN B 1 8 OD Tube Equal Tee 2ET2 B 1 4 OD Tube Equal Tee 4ET4 B 1 8 OD Tube Pressure Relief Valve This relief v...

Page 13: ...Generator The generator may be re configured so that the Parker Hannifin H and HMD range of hydrogen generators may be stacked safely on top of it To re configure the generator unlock the two retaini...

Page 14: ...or reaches the set temperature open the isolation valve located downstream of the generator Note The temperature may fluctuate up to 1 of the set temperature 6 The application instrument is now being...

Page 15: ...r leaks within the system Generator Check the mains supply cable for damage Generator Ensure there is ventilation clearance around generator Generator Replace the inlet and outlet filters Recommended...

Page 16: ...ck nut No 10 self tapping screw 120 V Fan Service Kit 606272526 606272526 Description Catalogue Number Technical Reference Contents 230 V Catalyst UHP 10ZA S 606272511 606272511 Catalyst assembly comp...

Page 17: ...Name Address Telephone Fax Contact Name Date of Commission Service Interval Months Hours Date Serviced By Comments Print Sign 6 4 000 12 8 000 18 12 000 24 16 000 30 20 000 36 24 000 42 28 000 48 32...

Page 18: ...sensor is open circuit Contact Parker Hannifin Indication Possible fault Condition Action Poor quality results downstream monitoring The generator is not powered Check the following 1 the generator i...

Page 19: ...supply cable is not damaged Contact Parker Hannifin Faulty controller or internal wiring fault Contact Parker Hannifin Indication Possible fault Condition Action Downstream monitoring Faulty cooling...

Page 20: ...r com NL The Netherlands Oldenzaal Tel 31 0 541 585 000 parker nl parker com NO Norway Asker Tel 47 66 75 34 00 parker norway parker com PL Poland Warsaw Tel 48 0 22 573 24 00 parker poland parker com...

Reviews: