DB Series Nitrogen Generator
DB5-20 Model
5. Communications
Ethernet Communications – DB 5-20 generators come standard with an Ethernet Modbus TCP/IP
interface located along the bottom edge of the controller. The port comes preconfigured in DHCP
mode. The MAC address is located on the front of the controller located inside the control box and
can be used to map the controller to a static IP address within the DHCP server. Modbus-RTU is
available using a separate gateway device.
5.1 MODBUS Register Map
Register Address
Description
00001
Unit is running
00002
Unit in Standby
00003
Unit left tower is generating
00004
Unit right tower is generating
00005
Unit tower pressures are equalizing
00006
Relay #1 contacts are closed
00007
Relay #2 contacts are closed
00008
Oxygen Probe range.
0 = 0.01 to 25.00 percent.
1 = 0 to 1000 ppm
00009
Unit has shut-down due to an alarm
00010
High Oxygen Alarm
00011
High Oxygen Warning
00012
Oxygen Sensor input fault alarm
00013
Flow sensor input fault alarm
00014
Pressure sensor input fault alarm
30001
Oxygen Purity
See register 00008 for sensor range
0 = 0 to 2500, Divide by 100 to calculate percentage
1 = 0 to 1000 ppm
30002
NST Flow – range depends on model
30003
NST Pressure – 0 to 200 psi
Summary of Contents for DB Series
Page 1: ...DB Series Nitrogen Generator Owner s Manual Models DB5 20...
Page 41: ...41 DB Series Nitrogen Generator DB5 20 Model DB 5 10 PPM PCT FD 9 Flow Schematics...
Page 42: ...42 DB Series Nitrogen Generator DB5 20 Model DB 15 20 PPM PCT FD...
Page 55: ...55 DB Series Nitrogen Generator DB5 20 Model Notes...
Page 56: ...56 DB Series Nitrogen Generator DB5 20 Model Notes...
Page 57: ...57 DB Series Nitrogen Generator DB5 20 Model Notes...