background image

 

 

                 1 

 

 

 

IMPORTANT:  

READ MANUAL COMPLETELY  BE F O R E   O P E R A T I N G   T H I S   D E V I C E  

Basic  delivery  technique  is  described.  Also,  this  manual  contains  instructions  on  periodically  required 
checks to be performed by the user.  These checks are necessary to insure the proper performance of 
this device and its safety features.  Retain this manual for future reference. 

 

 

W A R N I N G S   A N D   P R E C A U T I O N S  

 

These  warnings  and  precautions  are  to  help  you  to 
understand  how  to  safely  operate  the  MXR 
Flowmeter.    A  WARNING  alerts  you  to  a  possible 
hazard  to  people.    A  CAUTION  alerts  you  to  the 
possibility of equipment damage. 
 

WARNING:

    Do  not  use  this  device  for  the 

administration  of  general anesthesia or as a part of, 
or  in  conjunction  with,  a  general  anesthesia 
administration system. 
 

WARNING:

  Workers  are  exposed  to  N

2

O  during 

administration 

of 

N

2

O/O

conscious 

sedation 

analgesia. 

 

NIOSH 

has 

recommended  that 

exposures should be minimized.  Contact NIOSH (1-
800-35-NIOSH)  to  receive  NIOSH  Publications  on 

Control  of  Nitrous  Oxide  in  Dental  Operatories.  

Exposure  can  be  minimized  by  effective  controls.  
National Institute for Occupational Safety and Health 
(NIOSH)  publications  state  that  controls,  including 
System 

Maintenance, 

Ventilation 

and 

Work 

Practices  can  effectively  reduce  N

2

O  concentrations 

in  patient  procedures.    Your  flowmeter  accessory 
Porter  scavenger  system  is  an  important  part  of  the 
system of controls. 

 
WARNING:

  Porter Instrument equipment utilizes the 

cross+protection 

system.  The  flexible  hose  and 

connectors  that  connect  to  the  flowmeter  are 
diameter indexed; 3/8” O.D. for Nitrous Oxide and ½” 
for  Oxygen.  The 

cross+protection 

system  is 

designed  to  prevent  misconnection  of  Oxygen  and 
Nitrous  Oxide  piping.   

DO  NOT  ATTEMPT  TO 

CHANGE  THE  DIAMETERS  OR  CONNECTORS 
OF  THE  FLOWMETER!   

Tampering  with  the 

cross+protection

 system constitutes acceptance of 

liability  by  the  installer.  For  your  own  protection,  as 
well as that of the Doctor and the patients, use 3/8” 
O.D.  tubing  for  all  Nitrous  Oxide  lines  and  ½”  O.D. 
tubing for all Oxygen lines.  

To  assure  safe  operation  and  conformation  to  local 
fire  codes,  all  Porter  Instrument  flowmeter  systems 
are  designed  to  be  used  with  sedation  delivery 
systems  mounted  inside  walls  and  they  meet  or 
exceed  the  guidelines  established  by  the  National 
Fire  Protection  Association  for  Nonflammable 
Medical Gas Systems, NFPA 99. Copies of NFPA 99 
or  portions  thereof  may  be  obtained  by  writing  to:  

National  Fire  Protection  Association,  Batterymarch 
Park,  Quincy,  MA  02269-9904;  or  call  1-800-344-
3555 

 
WARNING:

  New or modified installations - 

properly 

connected gas pipelines are absolutely essential 
to  patient  safety.

    The  authorized  distributor  or 

contractor should provide written documentation that 
all gas pipelines are connected properly and that the 
system has been pressure tested prior to use.  While 
this  is  a  good  business  practice,  it  is  important  that 
the  user  verify  by  their  own  test,  independent of  the 
authorized  distributor  or  contractor,  that  all  gas 
pipelines  are  connected  correctly  prior  to  using  the 
system.   

The  ultimate  responsibility  of  assuring 

that lines are not crossed rests with the user. 
 
W AR N I N G :    

During any power outage, remember 

to  turn  OFF  the  flowmeter  and  manually  turn  OFF 
the  tank  valves.      With  centralized,  electrically 
powered  gas  systems,  if  gas  was  flowing  when  the 
power went out and the flowmeter is left ON, gas will 
be flowing when the power is restored. 
 

CAUTION:

  This device is for use by or on the order 

of  a  physician,  dentist  or  licensed  healthcare 
professional. 
 

CAUTION:

  Always  use  clean,  dry  medical  grade 

gases. 

 

Introduction 

of 

moisture 

or 

other 

contaminants into this device may result in defective 
operation. 
 

CAUTION: 

Do not attempt to repair, alter or calibrate 

this  device.    Unauthorized  repair,  alteration  or 
misuse of this device is likely to adversely affect the 
performance and will void the warranty. 
 

CAUTION:

  Never oil or grease any part of this 

system (minimize fire or explosion potential).

 

Summary of Contents for C3000

Page 1: ...Manufacturing Sarl Mailing Address Via privata Archimede 1 20094 Corsico Italy Phone 44 0 1271 313131 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration i...

Page 2: ...ection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire codes a...

Page 3: ...s set to 0 5 Concentration Control Knob controls the concentration of N2 O but does not change the O2 flow rate 70 Maximum N2 O delivery Check Monthly 6 Emergency Air Valve automatically provides the...

Page 4: ...itor O2 gas pressures at the beginning of each procedure to assure sufficient gas remains in the tank to complete the procedure SAFETY FEATURE CHECKS IMPORTANT These are safety features which you shou...

Page 5: ...ing bag is attached to the metal portion of the flowmeter Bag Tee Installation to Flowmeter Screw knurled seal down tight onto flowmeter making sure the rubber washer is inside the seal nut When tight...

Page 6: ...ation and post procedure 100 O2 will minimize nitrous exposure to the patient treatment area potential patient side effects of lethargy headache or nausea and any potential adverse effects of nitrous...

Page 7: ...empty 1 Turn ON N2 O cylinder 2 Replace with full cylinder With N2 O concentration set both flows vary proportionally with no change in flow setting O2 regulator is varying pipeline pressure Check O2...

Page 8: ...Similar to basic Model MXR design except different color coding for Oxygen white bands on flow tube and white buttons Flowmeter body is painted white and has a failsafe block cover that is also paint...

Reviews: