CONCENTRATION CONTROL KNOB VALVE 

– 

CHECK MONTHLY 

 

Set the Concentration Control Knob to 50%, and Flow 
Control  Knob  to  2  to  3  L/min.    The  ball  indicators  will 
be  at  about  the  same  height  (If  not,  refer  to 

Monthly

 

Calibration  Check

  section  for  the  procedure  on 

checking  the  accuracy  of  the  meter.).    Turn  the 
Concentration  Control  Knob  to  zero.    The  N

2

O  flow 

should  drop  to  zero.    You  essentially  perform  the 
check  at  the  end  of  every  procedure  when  you 
oxygenate the patient with 100% O

2

 

POWER OXYGEN FLUSH  
CHECK MONTHLY 
(O

2

 SUPPLY PRESSURE AT 50 PSIG) 

Disconnect the corrugated rubber tubing from the bag 
tee outlet (Figure 1, Item 11).  With both control knob 
valves OFF, depress the power (O

2

) flush button while 

blocking  the  flow  from  the  front  of  the  bag  tee.    For 
proper  operation,  the  gas  reservoir  bag  should  fill 
within  about  5  seconds.    Also  test  for  bag  /  rubber 
goods leak following steps 1 through 6. 

 

Quick Test to Check Reservoir Bag / Rubber 
Goods for Leaks 

 

1.  With  the  flowmeter,  bag  tee  and 

Porter

  rubber 

goods in place, remove the nosepiece and one of 
the  two  plastic  connectors  from  the  Porter  rubber 
goods.   

2.  With  the  other  plastic  connector,  join  the  two 

duplex hoses together making a closed system. 

3.  Taking care not to fill the bag too much (bag could 

burst),  open  the  oxygen  control  valve  until  the 
reservoir  bag  starts  to  over-

inflate  or  “balloon”, 

then close the valve. 

4.  Observe the reservoir bag for five minutes.  

5.  The  bag  should  stay  inflated.    If  so,  the  test  has 

been successful and there are no excessive leaks.  
If the bag does not stay inflated, the reservoir bag 
or rubber goods have an excessive leak.  Replace 
any  parts  that  leak  and  retest  until  results  are 
successful. 

6 .  

Disconnect  one  of  the  duplex  hoses  from  the 
plastic connector and re-install the nosepiece.

 

BAG TEE 

The  Bag  Tee  assembly  features  a  Non-rebreathing 
Valve  and  an  Emergency  Air  Intake  located  on  the 
Bag Tee (no carbon dioxide buildup).  The emergency 
air  intake  has  a  valve  that  allows  room  air  to  be 
inhaled  into  the  breathing  circuit  by  the  patient.    A 
breathing  bag  is  attached  to  the  metal  portion  of  the 
flowmeter.  

 Bag  Tee  Installation  to  Flowmeter:    Screw  knurled 
seal down tight onto flowmeter making sure the rubber 
washer is inside the seal nut.  When tight, the bag tee 
should not rotate. 

 

NON-REBREATHING VALVE  
CHECK MONTHLY  

With unit turned OFF, disconnect the corrugated rubber 
tubing  from  the  Rubber  Goods  and  breathe  into  the 
corrugated tubing connected to the bag tee.  You should 
not be able to fill the bag with exhalation gas.  If the bag 
fills,  the  system’s  Non-Rebreathing  Valve  is  not 
functioning properly and should be replaced. 

 

EMERGENCY AIR VALVE  
CHECK MONTHLY

 

With unit turned OFF, disconnect the corrugated rubber 
tubing  from  the  Rubber  Goods  and  draw  air  with  your 
mouth  through  the  corrugated  tubing  connected  to  the 
bag tee.  You should be able to draw ambient air through 
the Emergency Air Valve (the gas bag may have to fully 
collapse  first).      Air  going  through  the  valve  sounds 
different than normal gas flow. 

MONTHLY CALIBRATION CHECK 

The  MXR  Flowmeter  is  designed  to  maintain  its 
accuracy  and  performance  without  routine  user 
maintenance  being  required.    The  flowmeter  tubes  and 
ball  floats  are  very  resistant  to  accuracy  changes  over 
time  such  that  the  direct  readings  of  the  L/min  on  the 
scales  maintain  their  accuracy.    However,  the  user  can 
check the relative accuracies of the % concentration and 
total flow valve system by performing a simple check. 

A calibration check of the % concentration can be done 
by setting the % concentration knob to 50% and the flow 
control  knob  to  3  to  4  L/min.    Check  to  see  if  the  tube 
readings are within 0.5 L/min of each other.  Servicing is 
indicated if the readings are out of this tolerance. 

NOTE:  

Adequate and safe conscious sedation can be 

achieved even if the % concentration is outside of the 
listed  tolerance,  since  tube  scale  accuracies  are 
maintained.    However,  a  change  in  the  % 
concentration  calibration  is  an  indication  of  overall 
flowmeter  condition.    Porter  servicing  is  available, 
including  recalibration,  pressure  testing,  internal 
component  checking  and  replacement,  and  final 
factory 

testing 

by 

contacting 

your 

authorized 

distributor.    It  is  advisable, on  a  two  (2)  year cycle, to 
have  the  MXR  Flowmeter  factory  checked  and 
serviced. 

 

      

  

 

 

 

 

 

Summary of Contents for C3000

Page 1: ...Manufacturing Sarl Mailing Address Via privata Archimede 1 20094 Corsico Italy Phone 44 0 1271 313131 The Quality System for Porter Instrument is certified to ISO 13485 The scope of our registration i...

Page 2: ...ection as well as that of the Doctor and the patients use 3 8 O D tubing for all Nitrous Oxide lines and O D tubing for all Oxygen lines To assure safe operation and conformation to local fire codes a...

Page 3: ...s set to 0 5 Concentration Control Knob controls the concentration of N2 O but does not change the O2 flow rate 70 Maximum N2 O delivery Check Monthly 6 Emergency Air Valve automatically provides the...

Page 4: ...itor O2 gas pressures at the beginning of each procedure to assure sufficient gas remains in the tank to complete the procedure SAFETY FEATURE CHECKS IMPORTANT These are safety features which you shou...

Page 5: ...ing bag is attached to the metal portion of the flowmeter Bag Tee Installation to Flowmeter Screw knurled seal down tight onto flowmeter making sure the rubber washer is inside the seal nut When tight...

Page 6: ...ation and post procedure 100 O2 will minimize nitrous exposure to the patient treatment area potential patient side effects of lethargy headache or nausea and any potential adverse effects of nitrous...

Page 7: ...empty 1 Turn ON N2 O cylinder 2 Replace with full cylinder With N2 O concentration set both flows vary proportionally with no change in flow setting O2 regulator is varying pipeline pressure Check O2...

Page 8: ...Similar to basic Model MXR design except different color coding for Oxygen white bands on flow tube and white buttons Flowmeter body is painted white and has a failsafe block cover that is also paint...

Reviews: