background image

7

For assistance, call toll-free at 800-343-4048

Startup

Oxygen
Analyzer
Calibration

Operation/Calibration (Model HFX0-3)

Plug the IEC power cord into the power entry receptacle of the generator, and plug the opposite end
into a nearby wall outlet with earth ground protection.  (Note:

Note:

Note:

Note:

Note:  There is no power switch on the genera-

tor.  The oxygen analyzer is energized when the generator is plugged in.)

The inlet and outlet connections to the Balston HFX0-3 Nitrogen Generator must be checked for leaks
prior to system start-up.  After the system is properly installed and checked for leaks, the inlet gate
valve can be opened to introduce compressed air to the system.

During start-up and adjustment of the system, the nitrogen produced by the system will vary in purity.
If the application for the nitrogen is critically dependent on purity, the nitrogen produced during start-up
or adjustment should be vented.

If the outlet nitrogen flow is closed, the system will still consume compressed air.  The inlet air is simply
vented to atmosphere through the permeate ports of the membrane module.

CAUTION:  the oxygen analyzer will not provide accurate readings unless calibrated on a regular

CAUTION:  the oxygen analyzer will not provide accurate readings unless calibrated on a regular

CAUTION:  the oxygen analyzer will not provide accurate readings unless calibrated on a regular

CAUTION:  the oxygen analyzer will not provide accurate readings unless calibrated on a regular

CAUTION:  the oxygen analyzer will not provide accurate readings unless calibrated on a regular
basis.

basis.

basis.

basis.

basis.

The oxygen analyzer is calibrated prior to shipment;  however, Parker strongly recommends re-
calibrating the unit prior to initial start-up.  After the initial start-up, the analyzer should be calibrated on
a bi-weekly basis until a suitable schedule is determined, based upon the level of accuracy required by
the application.

There are two methods of calibrating the oxygen analyzer:  the two point method and the single point
method.  In the two point method,  the first point in the calibration range is set to zero using a zero gas
(zero percent oxygen), and the second point in the range is set to a known percentage of oxygen using
a span gas (known quantity of oxygen, per gas supplier) or compressed air (20.9% oxygen).  In the
single point method, only one point in the calibration range is set, using either span gas or compressed
air.  Maximum accuracy in oxygen concentration monitoring will be achieved if the oxygen concentra-
tion in the span gas is within the range of the expected oxygen concentration in the process stream
and the pressure of the gas closely approximates the pressure of the nitrogen gas (See Figure 3 for
calibration controls).

The procedure for the two point calibration method

two point calibration method

two point calibration method

two point calibration method

two point calibration method is as follows:

1

1

1

1

1 Throw the toggle switch on the back of the unit to the upright position, toward the calibration port.

2

2

2

2

2 Connect a tank of zero gas to the port. (Gas pressure should be approximately equal to operating

pressure.)

3

3

3

3

3 Allow the zero gas to flow through the unit until the reading on the oxygen concentration display

stabilizes.

4

4

4

4

4 Adjust the zero potentiometer until the oxygen concentration display reads zero.

5

5

5

5

5 Disconnect the zero gas from the calibration port, and connect a tank of span gas (operating

pressure) or a source of clean compressed air.

6

6

6

6

6 Allow the reading to stabilize, and adjust the span potentiometer until the reading on the oxygen

concentration display reads the known percent (span gas) or 20.9% (compressed air).

7

7

7

7

7 Disconnect the calibration gas and throw the toggle switch to the downward position to resume

sampling the nitrogen stream.

The procedure for the single point calibration

single point calibration

single point calibration

single point calibration

single point calibration method

method

method

method

method is as follows:

1

1

1

1

1 Connect a tank of span gas (operating pressure) or a source of clean compressed air to the calibra-

tion port.

2

2

2

2

2 Allow the gas or air to flow through the analyzer until the reading on the oxygen concentration

display stabilizes.

3

3

3

3

3 Adjust span potentiometer until the reading on the oxygen concentration display reads the known

percent (span gas) or 20.9% (compressed air).

4

4

4

4

4 Disconnect the calibration gas and throw the toggle switch to the downward position to resume

sampling the nitrogen stream.

(Note:  Parker recommends the use of a span gas with an oxygen content between 1% and 10%.)

!

Summary of Contents for Balston HFX-3

Page 1: ...separate air into a concentrated nitrogen output stream and an oxygen enriched permeate stream To monitor the purity of the nitrogen stream Parker has incorporated an oxygen analyzer into the design o...

Page 2: ...ure 1 Air separation takes place in the membrane module This module consists of bundles of hollow fiber membranes The compressed air enters the center bore of these fibers and travels the length of th...

Page 3: ...junction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times systems to assure proper nitrogen purity levels at all times systems to assure proper nitrogen...

Page 4: ...nt surrounding the nitrogen generator should also be adequately ventilated The environment surrounding the nitrogen generator should also be adequately ventilated The generator creates a 30 to 40 oxyg...

Page 5: ...sure to maintain the proper polarity black red The HFX0 3 Nitrogen Generator is preset at the factory for operation at 120 VAC The voltage setting for the generator is shown through a small window on...

Page 6: ...outside circuitry alarm relay contacts to outside circuitry alarm relay contacts to outside circuitry alarm relay contacts to outside circuitry The relay contacts are rated for 250 VAC 5 amps resistiv...

Page 7: ...gen using a span gas known quantity of oxygen per gas supplier or compressed air 20 9 oxygen In the single point method only one point in the calibration range is set using either span gas or compress...

Page 8: ...oxygen analyzer on a routine basis If the nitrogen purity level falls below the desired level readjust the flow control valve until the proper nitrogen purity level is reached Note If more than minor...

Page 9: ...vered from the system If the temperature difference is 10 F 20 F 5 C 11 C this equilibrium period may be as long as 60 minutes The inlet air tempera ture and more importantly inlet air dewpoint must n...

Page 10: ...vice schedule and replacement part numbers are outlined at the end of this section The calibra tion procedure for the oxygen analyzer is outlined on page 6 The analyzer should be calibrated on a regul...

Page 11: ...en servicing This equipment has fuses in both neutral and phase lines Use care when servicing Occasionally one or both of the fuses P N 13221 in the generator may burn out The fuses are located in the...

Page 12: ...ve 95 F 35 C the longevity of the membrane will be reduced and the warranty will be void 6 6 6 6 6 The drain lines from the first two stages of filtration should be piped away to an appropriate collec...

Page 13: ...58 F 50 C Particles 0 01 m None None Suspended Liquids None None Commercially Sterile Yes Yes Max Operating Pressure 145 psig 10 barg 145 psig 10 barg Maximum Pressure Drop 10 psi 0 7 bar 10 psi 0 7 b...

Page 14: ...d range during calibration Replace sensor P N 72695 Operating Conditions To arrange for system service contact the Technical Services Department at 1 800 343 4048 8AM to 5PM Eastern Time For location...

Page 15: ...cription Description Description Description Caution refer to accompanying documents for explanation Refer the caution warning note indicated for explantion Caution risk of electric shock 3 3 3 3 3 Ex...

Page 16: ...rhill MA 01835 0723 Tel 978 858 0505 Fax 978 556 7501 Parker Filtration Separation Parker Filtration Separation Parker Filtration Separation Parker Filtration Separation Parker Filtration Separation F...

Reviews: