Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

5

1-800-343-4048

All maintenance activities should be performed by suitable personnel using reasonable 
care. Safety risks that may affect the service personnel are identified with protective 
measures that apply. The service personnel will verify the safe state of the analyzer 
after maintenance is completed.

The only routine maintenance required by the Balston 72-730 and 72-O2730NA Oxygen 
Analyzers is timely calibration and annual sensor replacement.  Calibration details and rec-
ommended calibration scheduling are outlined in the “Calibration” section of this manual.  If 
necessary, the Balston 72-730 and 72-O2730NA Oxygen Analyzer may be wiped clean with a 
dry cloth on an as needed basis.  

Do not use water, aerosols, or other cleaning agents to 

clean the unit.

Note:

  There are no customer serviceable parts inside the metal cover of the analyzer.

Disconnect analyzer from power supply before replacing galvanic cell sensor.

The galvanic cell sensor degrades over time and should be replaced on an annual basis              
(P/N 72695).  The only tools needed for this replacement are a small screwdriver and wire 
strippers.  The procedure for changing the sensor is outlined below and takes approximately 
10 minutes.

Remove sensor cover and disconnect old cell wires from connectors by pressing the 
screwdriver into the hole next to the connector and pulling wires out.

Strip replacement sensor connecting wires to 1/4" to 3/8" (6mm to 9mm) using wire 
strippers.  Install the sensor into the sensor holder on the left side panel of the analyzer 
(see Figure 3) by pushing firmly into place.

3

 

Insert small screwdriver into the hole adjacent to the wire connection point (see Figure 3) 
and press to open connector.

4

 

Slide the stripped wire end into the appropriate (red or black) connection port until it 
“bottoms out”.  The red wire goes to the “+” connector, and the black wire goes to the 
“-” connector.

5

 

Remove the screwdriver to clamp the wire into the connection port.  Pull the wire gently 

Maintenance

Galvanic Cell 

Replacement

The customer is responsible for the circuitry utilizing these relay outputs and should use good 
engineering safety practices in the design of this circuitry.

Strip all connecting wires to 1/4" to 3/8" (6mm to 9mm) using wire strippers.  

Insert small screwdriver into the hole adjacent to the wire connection point and press to 
open connector.

Slide the stripped wire end into the connection port until it “bottoms out”.

Remove the screwdriver to clamp the wire into the connection port.  Pull the wire gently to 
test integrity of the connection.  Repeat this procedure from step 2 if the wires release easily.

Figure 6 - Sample Schematic for Wiring Alarm Relay Contacts, Right Side View

Summary of Contents for Balston 72-730

Page 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Page 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Page 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Page 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Page 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Page 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Page 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Page 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...

Reviews: