Parker Balston 72-730 Installation, Operation And Maintenance Manual Download Page 2

Bulletin TI-72730J 

Installation, Operation and Maintenance Manual

 

72-730 and 72-O730NA Oxygen Analyzer

www.labgasgenerators.com

2

1-800-343-4048

Alarm Controls -

  The alarm controls are located on the front panel of the analyzer.  The switch 

on the far right enables the audible alarm.  When enabled, the audible alarm will sound if the 
oxygen concentration in the process stream exceeds the alarm set points set by the user.  The 
alarm set switch is located to the left of the audible alarm control switch.  The two potentiom-
eters used to set alarm trigger points are located to the left of the alarm set switch.  The  LED’s 
above and below the alarm set switch give a visual indication that the oxygen concentration is 
beyond the specified range.

Oxygen Concentration Display -

  The oxygen concentration LED display shows oxygen con-

centration, in percent, to the nearest 0.1%.

Calibration Controls -

  The calibration controls are located to the left of the oxygen concentra-

tion display.  The zero potentiometer is used to zero the instrument when a zero gas (containing 
no oxygen) is introduced.  The span potentiometer is used to set the 72-730 and 72-O2730NA 
to the specified concentration of oxygen in the span gas.  

Controls (cont’d)

Figure 4- Recommended Calibration Gas Valving

Figure 2 - Mounting Dimensions

All installation, operation and maintenance procedures for the Balston 72-730 and              
72-O2730NA

 

Oxygen Analyzer should be performed by suitable personnel using reason-

able care.

The Balston 72-730 and 72-O2730NA Oxygen Analyzers should be located indoors and as close 
to the process stream as possible to ensure accurate readings.  

Do not install the oxygen ana-

lyzer outdoors.  

The location should be clean, dry, and between 59°F and 95°F (15°C - 35°C).  

If the unit will be wall-mounted, use the mounting holes on the back of the oxygen analyzer.  
Mounting dimensions for the units are shown in Figure 2.  Any mounting hardware supplied by 
the customer should be adequately sized to support the weight (5 lbs./2.2 kg) of the oxygen 
analyzer in its mounted location.  The unit should be mounted to the wall using National Electri-
cal Code (NEC) and local building code guidelines.  Allow a minimum of 2 in. (5 cm) on all sides 
of the analyzer. 

The inlet port to the 72-730 and 72-O2730NA is a 1/8" single ferrule fitting and is located on the 
left side of the unit, as indicated in Figure 3.  The flow rate through the oxygen analyzer is con-
trolled by a fixed orifice located inside the unit.  The inlet pressure to the analyzer must be held 
between 2 psig and 140 psig (0.14 barg and 10 barg), to yield an accurate reading.          (

Note:

  

Parker recommends setting the pressure of the calibration gases as close to the operating pres-
sure of the system as possible to ensure optimal accuracy in oxygen concentration readings.)

The sample line and fittings should be brass or stainless steel to prevent contamination of the 
process stream sample.   A schematic for the recommended process stream and calibration gas 
valving is shown in Figure 4.

Installation

Sample lines

Figure 3 - Left Side View

With Cover Removed

Summary of Contents for Balston 72-730

Page 1: ...chemicals or packaging food for extended shelf life an oxygen moni tor and or trace oxygen analyzer should be used in conjunction with safety interlocks and or alarm systems to assure proper nitrogen purity levels at all times The Balston Oxygen Analyzer is a self contained unit which may be wall mounted bench mounted or ordered as an integral accessory to any Nitrogen Generation System The unit i...

Page 2: ...le personnel using reason able care The Balston 72 730 and 72 O2730NA Oxygen Analyzers should be located indoors and as close to the process stream as possible to ensure accurate readings Do not install the oxygen ana lyzer outdoors The location should be clean dry and between 59 F and 95 F 15 C 35 C If the unit will be wall mounted use the mounting holes on the back of the oxygen analyzer Mountin...

Page 3: ...g the opposite end into a wall outlet with earth ground protection The North American cordset is rated for 125V 10 amperes 18 AWG SJT with ferrite bead Do not replace it with an inadequately rated cordset The analyzer is powered on by plugging into electrical supply There is no power switch on this product If the voltage selector displays an input power voltage different from the local power suppl...

Page 4: ...the oxygen concentra tion display stabilizes 5 Adjust the span potentiometer until the reading on the oxygen concentration display reads the known percent span gas or 20 9 compressed air 6 Close the valve from the calibration gas line and re open the valve between the process stream and the analyzer Note Parker recommends the use of a span gas with an oxygen content between 1 and 10 The high and l...

Page 5: ...approximately 10 minutes 1 Remove sensor cover and disconnect old cell wires from connectors by pressing the screwdriver into the hole next to the connector and pulling wires out 2 Strip replacement sensor connecting wires to 1 4 to 3 8 6mm to 9mm using wire strippers Install the sensor into the sensor holder on the left side panel of the analyzer see Figure 3 by pushing firmly into place 3 Insert...

Page 6: ...er requirement 120 VAC 230 VAC 50 60 Hz 10 A 120 VAC 05 A 230 VAC Fuse Type Type T Slow Blo Fuse 250 V 0 10 amp 5 x 20 mm Physical dimensions 11 h x 5 w x 4 d 28cm x 13cm x 11cm Shipping weight 6 5 lbs 3 kg Note Main supply line voltage must be within 10 of nominal rated voltage for the generator System Specifications 2 3 Fuse Replacement to test integrity of the connection Repeat this procedure f...

Page 7: ...ice is required please record the following DATE IN SERVICE SERIAL NO Please have the serial number available when calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or is defective within 12 months from the date of purchase when this product is operated ...

Page 8: ... 482 6296 www parker com gsfe FNSdhFNS_TSG parker com Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 4087 Walden Ave Lancaster NY 14086 Tel 978 858 0505 Fax 978 478 2501 www labgasgenerators com Symbol Description Caution refer to accompanying documents for explanation Refer to the caution warning note indicated for explanation Caution risk of electric shock 3 Copyri...

Reviews: