Bulletin TI-NITROFLOW TG2-G
Installation, Operation and Maintenance Manual
TI-NITROFLOWTG2
www.labgasgenerators.com
15
1-800-343-4048
Fuse Replacement
10,000 Hour Service
Instructions
12
Place dust seal over the backup tube. (Figures 33, 34)
Caution! The backup tube and dust seal must meet at the bottom of the housing
in the six o’clock position.
13
Install the stationary scroll housing onto the orbiting scroll housing and reassemble the unit.
14
Tighten the pulley bolt (M8) to 265 in -lbs, using a torque wrench.
15
Tighten the six housing nuts to 265 in-lbs, using torque wrench.
There are three fuses located in the compressor cabinet (see Figure 35).
Before servicing the fuses, turn
the generator off and disconnect the power cord from the power supply.
To access the fuses, remove the upper compressor access panel (see Figure 1), and locate the fuse blocks.
Both lines 1 and 2 are fused separately. Remove both line fuses with a small screwdriver. Replace either
one or both fuses as required.
To maintain safety, use only fuses of the size and type specified in the
Specifications section of this manual.
These instructions describe the replacing of the two coalescing pre-filters, two carbon modules, and two
particulate filters in the NitroFlow TG2 Generator System Series.
Generator must be set to “OFF” and pressure gage must read zero before filter replacement.
Figure 35
Reset
Button
DX
Water
Separator
BX
Figure 36
Figure 33
without dust seal
Figure 34
with dust seal
Coalescing Filters
1
The replacement elements are B100-18-DXE (first stage) and BY100-18-BXE (second stage).
It is very important to install the proper grade filter into the proper housing. Take extra
care to install in the proper sequence (see Figure 36). From the rear, DX on left, BX on right.
2
Twist the bowl loose from the head of the assembly, counter clockwise, and remove carefully
so as not to damage or lose the o-ring (see Figure 37).
3
Unscrew the element retainer at the base of the element, and replace the spent element with
a new one (see Figure 38).
4
Reverse the steps to reassemble. Make sure the o-ring is sealed in the groove before
reinstalling the bowl.
Figure 37
Figure 38
Generator must be set to
“OFF” and pressure gage
must read zero before filter
replacement.