Bulletin TI-H2PEMPD-E
Installation, Operation and Maintenance Manual
Hydrogen Gas Generator H2 PEMPD series
23
1-800-343-4048
www.labgasgenerators.com
Generator/Cell HIGH
pressure
Off
Contact tech support/
PSU. PCB, Transducer
fault
No cell pressure signal
Off
Reset/replace, check
connection
No output pressure
signal
Off
Reset/replace, check
connection
Faulty, no signal
On
Reset/ check sensor
connection
Cell separator water
level HIGH
Off
Check/replace sensor or
connection
Calibration file out of
date
On
Calibrate conductivity
sensor
Software timeout
Off
Reset/contact tech
support
Pump fault
– no cell
water circulation
Off
Check/replace pump or
connection
Power supply fan not
rotating
Off
Reset/replace the fan
Case cooling fan not
rotating
On
Reset/replace the
external fan
Reservoir level sensor
faulty
Off
Reset/check connector,
replace sensor
Purifier temp sensor
faulty
Off
Reset/check connector.
Replace heater
OVER PRESSURE
ACT 100%
PRESS XDUCER 1
ACT 100%
PRESS XDUCER 2
ACT 100%
COND XDUCER
ACT 100%
SEPARATOR BLOCK
ACT 100%
CALIBRATION
ACT 100%
WATCHDOG
ACT 100%
WATER PUMP
ACT 100%
PSU FAN FAIL
ACT 100%
CASE FAN FAIL
ACT 100%
O2 FLOAT CONNECT
ACT 100%
TEMP XDUCER
ACT 100%