5.3.5 Replacing the Fan Guards
Remove the 4 retaining screws from each fan guard (figure 18, item 7). Fit the replacement fan guards and secure using the
previously removed screws.
5.3.6 Filling the water bottle
The use of any water, other than deionized water (ASTM II, >1 MΩ, <1µS, filtered to <100µm within this generator
will damage and reduce the life time of the hydrogen cell.
Remove the top front cover and the water bottle cap. Fill the water bottle using fresh deionized water to a level approximately 15mm
below the upper lip of the neck of the bottle.
If the generator is powered during refilling, an audible and visual indication will be given by the generator when the correct level is
reached, and the LCD
will display “Water Full” message. Once full replace the water bottle cap and top front cover.
Note:
If the water has been changed due to high conductivity, the deionizer cartridge must also be changed.