background image

Optimize Scan Parameters 

 
 
1) Click on Input configuration icon 

 

2) Select channels of interest, show at least trace and retrace of topography 

3) Change the scan size to appropriate value (5µm or 50µm max depending on XY scanner) 

 
A good image is achieved when trace and retrace show the same topography and are stable 
(repeatedly show the same topography and there are no random spikes or other instabilities).  
Trace and retrace may not overlap perfectly but if the shapes of the curves are identical, then the 
image will generally be good. To improve the tracking, adjust the following: 
 

4) 

Z-servo gain

 - Increasing the gain will help the tip track the sample better.  However, if the 

gain is too high, you will see high-frequency noise.  Increase/decrease the gain until the best 
image is obtained.  In most cases, the optimal value will be from 1-5. 

5) 

Scan rate - 

Decrease the scan rate to 0.5 Hz or 0.25 Hz.  This effectively reduces the speed at 

which the tip moves, allowing the system more time to adjust to sudden changes in height.  Be 
careful, as this will also proportionately increase the time needed to take images. 

6) 

Set Point - 

Adjusting the Set Point up/down will change the distance the tip hovers above the 

sample surface when scanning and can improve resolution of features/surface tracking. This can 
be done manually by changing the number in the ‘Set Point’ box or by adjusting the red bar in 
the frequency sweep window.  

7) 

Drive % - 

Adjust the Drive% by a few tenths of a percent at a time to see if your image 

quality improves. This will increase the force with which the tip oscillates.  Only adjust this 
setting after exhausting the other options as increasing the drive can deteriorate the tip life. 

 

 

 

Getting a good image can take time.

  It is a matter of iterating between adjusting the gain, set 

point and drive. If you are unable to get a good image, there are three additional options:   
 

1) Hit the “NCM ASetup” button at the bottom of the scan parameters window.  This will run an 
automatic routine to try and find a good set of scan parameters.  The frequency window will 
open automatically.  Hit OK to close this window.  

2) If you still don’t see your surface, retract the tip from the sample, increase the initial drive %, 
and re-approach the sample. 

3) If you still don’t see your surface, it may be that the tip you are using is damaged.  Raise the 
optical microscope, then raise the Z-stage well away from your surface.  Remove the head.  Put 
the tip in the ‘Bad Probes” box and select a new probe from the ‘Good Probes’ box. 

Summary of Contents for XE-100

Page 1: ...ser will move on the quadrant photodetector This is relayed to a motor controlling the tip which adjusts the tip s vertical position until the laser is aimed in the center of the quadrant photodetecto...

Page 2: ...ting tip and place it in the good tips box Ensure new tip is rested in place and that bearings fit into tip slots 4 Replace head and push flaps away from you until slightly tight This secures the head...

Page 3: ...n Look at where the light from the microscope is shining on the tip assembly as this will help you to center the microscope Position laser on cantilever If you don t see the laser reflected off the ca...

Page 4: ...e appropriate value If after several tries you still can t get the sum high enough remove the head and check mirror 2 s position visually it should be roughly parallel to the base Adjust mirror 2 knob...

Page 5: ...o 1 00 Cantilever General 4 Click OK 5 Turn Head back on in software Frequency Sweep Window will open 6 The software will automatically detect the cantilever s resonance peak and select a driving freq...

Page 6: ...B and C D are still less than 0 3 V 9 Shut acoustic enclosure 10 Click Approach button in Z stage control When the approach stops the light in the Z stage Control window will stop blinking Ensure tha...

Page 7: ...will change the distance the tip hovers above the sample surface when scanning and can improve resolution of features surface tracking This can be done manually by changing the number in the Set Poin...

Page 8: ...h XEI You can send images directly to the XEI data analysis software from the Buffer in XEP by right clicking on them Or you can open them from the spmdata or User Files folders Flattening your sample...

Page 9: ...er as needed 5 Turn off laser switch unplug head and remove head 5a If the tip you used was good leave it on the AFM Otherwise remove head and place tip in the Bad Probes box 6 Remove sample from stag...

Page 10: ...ion on center of photodetector A B 3 5V A B and C D 0 3V 6 Select appropriate mode scanner and frequency 7 Approach surface i Get optically close several mm ii Focus microscope 50 100um from surface m...

Reviews: