background image

Mode, Scanner, and Frequency Selection 

 
 
1) Turn off Head in software 
 
 
 
2) Click on Parts Select 
 
 
 
3) Select proper mode of operation and scanner ranges 

Head Mode: NC-AFM - 

ACTA tips should only be used for NC-AFM (non-contact mode)

 

XY Voltage Mode: High (50um x 50um) or Low (5umx5um) 
Z Voltage Mode: High (12 um) or Low (1.7 um) 
Z Scanner Range: Set to 1.00  
Cantilever: General 

4) Click OK 

5) Turn Head back on in software, Frequency Sweep Window will open. 

6) The software will automatically detect the cantilever’s resonance peak and select a driving 
frequency just above that peak.  

- If a very weak peak appears (peak is on the order of the noise) or if there is strong non-
ideal behavior like a double peak, replace the tip.  An additional indication of a bad 
cantilever is a Drive % at/near 100. This is a rare occurrence under normal operation. 

7) Click OK 

 

 

Bad (low amplitude, 100% drive)                        Good 

 

Summary of Contents for XE-100

Page 1: ...ser will move on the quadrant photodetector This is relayed to a motor controlling the tip which adjusts the tip s vertical position until the laser is aimed in the center of the quadrant photodetecto...

Page 2: ...ting tip and place it in the good tips box Ensure new tip is rested in place and that bearings fit into tip slots 4 Replace head and push flaps away from you until slightly tight This secures the head...

Page 3: ...n Look at where the light from the microscope is shining on the tip assembly as this will help you to center the microscope Position laser on cantilever If you don t see the laser reflected off the ca...

Page 4: ...e appropriate value If after several tries you still can t get the sum high enough remove the head and check mirror 2 s position visually it should be roughly parallel to the base Adjust mirror 2 knob...

Page 5: ...o 1 00 Cantilever General 4 Click OK 5 Turn Head back on in software Frequency Sweep Window will open 6 The software will automatically detect the cantilever s resonance peak and select a driving freq...

Page 6: ...B and C D are still less than 0 3 V 9 Shut acoustic enclosure 10 Click Approach button in Z stage control When the approach stops the light in the Z stage Control window will stop blinking Ensure tha...

Page 7: ...will change the distance the tip hovers above the sample surface when scanning and can improve resolution of features surface tracking This can be done manually by changing the number in the Set Poin...

Page 8: ...h XEI You can send images directly to the XEI data analysis software from the Buffer in XEP by right clicking on them Or you can open them from the spmdata or User Files folders Flattening your sample...

Page 9: ...er as needed 5 Turn off laser switch unplug head and remove head 5a If the tip you used was good leave it on the AFM Otherwise remove head and place tip in the Bad Probes box 6 Remove sample from stag...

Page 10: ...ion on center of photodetector A B 3 5V A B and C D 0 3V 6 Select appropriate mode scanner and frequency 7 Approach surface i Get optically close several mm ii Focus microscope 50 100um from surface m...

Reviews: