F.  CLEANING YOUR PARI REUSABLE 

NEBULIZER

1.  Disconnect the tubing from the air intake on the 

bottom of  the nebulizer cup.

2.  If  there is any moisture (condensation) in the tubing, 

let the compressor run with only the tubing attached 
for 2-3 minutes. The warm air from the compressor 
will dry out the tubing. Alternatively, you can dry 
the tubing by removing it from the compressor and 
hanging it up with the “ends down” to allow any 
moisture (condensation) to drain out and evaporate. 
Use a clean damp cloth to wipe the exterior of  
the tubing.

3.  Disassemble reusable nebulizer parts as illustrated 

on page 3.

4.  Wash all parts with warm water and liquid dish soap. 

DO NOT WASH THE TUBING.

5.  Rinse the reusable nebulizer parts thoroughly with 

warm water and shake out water.

6.  Air dry or hand dry nebulizer parts on a clean, lint-free 

cloth. Reassemble nebulizer when parts are dry 
and store.

7.  You can also wash all parts of  the reusable nebulizer 

on the top shelf  of  a dishwasher. DO NOT WASH 
TUBING. Place the reusable nebulizer parts in a 
dishwasher basket, then place the basket on the top 
rack of  the dishwasher. Remove and dry the parts 
when the dishwasher cycle is complete.

WARNING

TO REDUCE RISK OF INFECTION:

1.  Nebulizer for single patient use (one person) only. Do 

not share your nebulizer with other patients.

2.  You must regularly disinfect or sterilize the nebulizer 

between treatments. Failure to do so could lead to 
serious or fatal illness.

REMEMBER

 REPLACE YOUR REUSABLE NEBULIZER CUP EVERY SIX MONTHS

* Lester, Mary K. Nebulizer Use and Maintenance by Cystic Fibrosis Patients: A Survey Study. Respiratory Care 2004; 49(12): 1504-1508.

Advantages of Replacing Your Nebulizer Cup: 

•  Consistent medication delivery to the lungs

 

•  Fast and effi cient treatments   

 

•  Bacterial contamination can be reduced*

F

Questions? 1.800.FAST.NEB (327.8632)  |  PARI.com/VIOS

EN-5

E. TIMESTRIP

®

It is very important to replace your PARI reusable nebulizer 
every 6 months for the most effective treatments. 
Timestrip

®

 is a visual indicator that makes it easy for you to 

remember when to replace your nebulizer. 

Activating Timestrip

®

:

 Firmly squeeze the PARI bubble 

between your fi nger and thumb. You may need to squeeze 
it more than once to ensure all the liquid has been emptied 
from the bubble.

Verify Activation:

 A red line will appear in the small circle of  

the Timestrip

®

 within 1 to 2 minutes, indicating activation. 

If  you do not see the red line after 3 minutes, press the 
bubble again.

Attach to compressor:

 Remove the adhesive tab on the back 

of  the Timestrip

®

 and attach the Timestrip

®

 to the front of  

your compressor as indicated in Section C. IMPORTANT – 
DO NOT ATTACH Timestrip

®

 to your NEBULIZER CUP.

Replace Nebulizer:

 When you see the word NOW appear 

in the window (Fig. E), you know it’s time for a new PARI 
reusable nebulizer! Contact your home healthcare provider 
or pharmacy to order your new PARI reusable nebulizer.

WARNING

The Timestrip may present a choking hazard to small 
children – adult supervision is required.

Timestrip

®

 does not monitor the condition of the PARI reusable nebulizer. Timestrip

®

 

only records the passage of time after it is activated. If activated at the time of fi rst 
using the PARI reusable nebulizer then it will serve as a reminder that the PARI reusable 
nebulizer can be replaced.   

The liquid inside the Timestrip

®

 is not toxic.  If tampered with, the Timestrip

®

 may leak 

and will not function properly. If the liquid comes into contact with the skin wash with 
soap and water.  If the liquid comes into contact with eyes rinse immediately with water 
for 15 minutes. If irritation develops, seek medical attention.

Timestrip

®

 is a registered trademark of Timestrip Plc 

E

Vios IFU- Sprint EN 0029.indd   6

Vios IFU- Sprint EN 0029.indd   6

6/16/10   11:47:14 PM

6/16/10   11:47:14 PM

Summary of Contents for Vios

Page 1: ...T READ INSTRUCTIONS AND SAFETY PRECAUTIONS BEFORE USE READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE USE THE PRODUCT ONLY FOR ITS INTENDED USE AS DESCRIBED IN THESE INSTRUCTIONS FOR USE IMPROPER USE CA...

Page 2: ...5 This product contains small parts that may present a choking hazard to small children The tubing also presents a strangulation hazard ALWAYS USE CLOSE ADULT SUPERVISION WHEN ADMINISTERING A TREATMEN...

Page 3: ...ly in the outlet reverse the plug 5 Never block the air vents of this compressor or place it on a soft surface such as a bed or couch where the air vents may be blocked Keep the air vents free of lint...

Page 4: ...er 2 included PARI Reusable Nebulizer Holder Clip Air Outlet Air Filter Power Switch Timestrip Timestrip Location Inspiratory Valve Cap Nebulizer Outlet Mouthpiece with Expiratory Valve Nebulizer Top...

Page 5: ...bottom of PARI reusable nebulizer Fig D 7 StartingTreatment Sit in a relaxed upright position and turn the compressor ON 8 Nebulizer Position Place the mouthpiece on top of tongue and close lips aroun...

Page 6: ...e reduced F Questions 1 800 FAST NEB 327 8632 PARI com VIOS EN 5 E TIMESTRIP It is very important to replace your PARI reusable nebulizer every 6 months for the most effective treatments Timestrip is...

Page 7: ...ecomes clogged A good rule to follow is to replace your filter every 6 months at the same time as your PARI reusable nebulizer Removing First make sure the Vios power switch is turned off Using a coin...

Page 8: ...nning NOTE WHEN SWITCH IS IN THE OFF POSITION AND THE COMPRESSOR IS NOT RUNNING THERE WILL STILL BE ELECTRICAL POTENTIAL INSIDE THE COMPRESSOR HOUSING TO REMOVE POWER FROM THE COMPRESSOR THE POWER PLU...

Page 9: ...TIES ON THE Vios Aerosol Delivery System AND ITS PARTS INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED TO SIXTY 60 MONTHS FROM THE DATE OF FIRST RETAIL...

Page 10: ...or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospi...

Page 11: ...d from the equation applicable to the frequency of the transmitter Recommended separation distance P V d 1 P E d 1 80 MHz to 800 MHz P E d 1 7 800 MHz to 2 5 GHz where P is the maximum output power ra...

Page 12: ...ES DE SEGURIDAD ANTES DE USAR EL DISPOSITIVO LEA TODAS LAS ADVERTENCIAS E INSTRUCCIONES ANTES DE UTILIZAR EL EQUIPO USE EL PRODUCTO S LO PARA EL PROP SITO PARA EL QUE FUE DISE ADO QUE SE DESCRIBE EN E...

Page 13: ...o 5 E l producto contiene piezas peque as que pueden presentar un riesgo de asfixia para ni os peque os La entubaci n adem s presenta un riesgo de estrangulaci n SIEMPRE CUENTE CO N LA SUPERVISI N DE...

Page 14: ...macorriente d lo vuelta 5 Nunca obstruya las ventilaciones de aire de este compresor ni lo coloque en una superficie blanda como una cama o sof donde se puedan obstruir las ventilaciones Mantenga las...

Page 15: ...lizador reutilizable PARI Salida de aire Filtro de aire Interruptor el ctrico Timestrip Ubicaci n sugerida de Timestrip Tapa de la v lvula de inspiraci n Salida del nebulizador Boquilla con v lvula de...

Page 16: ...la parte inferior del nebulizador reutilizable PARI Fig D 7 Inicie el tratamiento Si ntese en una posici n relajada y recta y coloque el compresor en ON 8 Posicione el nebulizador Coloque la boquilla...

Page 17: ...dedo y el pulgar Es posible que deba apretarla m s de una vez para asegurarse de que haya salido todo el l quido de la burbuja Verifique la activaci n Aparecer una l nea roja en el c rculo peque o de...

Page 18: ...obstruye Una buena norma es reemplazar el filtro cada 6 meses al mismo tiempo que su nebulizador reutilizable PARI Extracci n En primer lugar aseg rese de que el interruptor el ctrico del Vios est apa...

Page 19: ...BOLOS Atenci n Lea todas las Instrucciones de uso antes de usarlo Equipo de aislamiento cerrado Clase II Protecci n tipo B contra choque el ctrico No lo elimine como desechos no clasificados Consulte...

Page 20: ...por Aerosol Vios Y SUS PIEZAS INCLUIDAS LAS GARANT AS IMPL CITAS DE COMERCIABILIDAD Y ADECUACI N PARA UN PROP SITO PARTICULAR SE LIMITAN A SESENTA 60 MESES A PARTIR DE LA FECHA DE LA PRIMERA COMPRA E...

Reviews: