4.4 Replacing the air filter
The air filter must be checked at regular intervals (after every
10th application). If it is discoloured brown or grey, or if it is
damp or clogged, it must be replaced.
In all cases, it must be replaced after 200 operating hours (ap-
prox. 1 year).
The air filter cannot be cleaned and then reused!
Take the air filter out:
• Pull the filter holder out of the
compressor. Use a small
screwdriver, for example, to
prise the filter holder carefully
out of the compressor.
Replace the air filter:
• Pull the old air filter out of the filter holder and fit the new fil-
ter in its place.
NOTE!
Only use air filters provided by the manufacturer or
dealer for operating your compressor. If air filters that are
not designed for the compressor are used, the compressor
may be damaged.
• Insert the filter holder back in the compressor.
4.5 Inspecting
Inspect all product components after each cleaning and disin-
fection. Replace any broken, misshapen or seriously dis-
coloured parts.
– 33 –