4.4 Disinfection
After cleaning, disinfect the dismantled nebuliser and accessor
ies used (except for the connection tubing) (effective disinfec
tion is only possible if the nebuliser has been cleaned).
Information when using masks with rubber bands:
The rubber band can only be cleaned as it might be dam
aged by the high temperatures used during disinfection.
CAUTION
A damp environment may encourage the growth of bacteria.
Therefore, remove all parts from the pot or disinfector as soon
as disinfection has finished. Dry the parts. The
risk of infec
tion
is reduced when the parts are dried completely.
In boiling water
• Place all the individual parts in boiling water for at least
5 minutes.
Use a clean pot and fresh drinking water.
If your water is very hard, scale may be deposited on the
components as a milky coating. This can be prevented by
using softened water.
NOTICE
Plastic will melt if it comes into contact with the hot base of the
pot. Therefore, make sure there is plenty of water in the pot.
This way you will avoid
damaging the components
.
– 52 –
Summary of Contents for MHL T
Page 1: ...LC SPRINT XLent 2014 PARI GmbH Spezialisten f r effektive Inhalation 023D1030 A 07 14...
Page 12: ...Nasa te na nebuliz tor p ipojovac trubici Pou v n s n ustkem Nasa te na nebuliz tor n ustek 12...
Page 34: ...34...
Page 64: ...64...
Page 128: ...128...
Page 160: ...160...
Page 192: ...192...
Page 193: ......