MAGNAIR questions? Contact the Sunovion Customer Service line (1-888-394-7377).
22
Instructions for Use
For use with LONHALA
TM
(glycopyrrolate)
Inhalation Solution
Fault
Possible Causes / Resolutions
6
MAGNAIR will not switch off
automatically.
Turn MAGNAIR off by pressing the On/Off button.
7
Aerosol mist continuously
escapes in large volumes
from the opening slots in the
Medication cap.
Improper Handset assembly. Check that the Handset
has been assembled correctly and the Blue valve flap
is pushed down.
8
The Medication cap will not
close.
LONHALA vial is not seated
properly. Press the LONHALA
vial base toward the cap
until you hear it snap
into place.
9
The LONHALA vial has been
inserted incorrectly.
If the LONHALA vial has been inserted incorrectly, the
following steps must be carried out:
1. Open the Medication cap.
2. If the LONHALA vial has been opened, throw away the
LONHALA vial, rinse and dry the Handset parts, then
insert a new vial.
3. If the LONHALA vial has not been opened, place the
Medication cap back on the Handset and turn the
Medication cap containing your LONHALA vial in a
clockwise direction until it clicks, then restart your
treatment by pressing the On/Off button.
10
Increasing or long inhalation
time.
1
Treatment time may vary
up to 15 minutes if the Aerosol
head has become clogged.
The Aerosol head may be clogged. Clean the Aerosol
head by soaking it in soapy water for 5 minutes and rinse
both sides well
(
Section F
)
.
11
Top of Handset body has
become detached from bottom
of Handset body.
Line up the hinge part on the
top of the Handset body with
the hinge portion of the bottom
of the Handset body. Press firmly
until the parts snap together. You
should hear a Click.
Click
1
If the fault cannot be eliminated after following these steps, call the Sunovion Customer Service line
(1-888-394-7377) immediately.
G. TROUBLESHOOTING (cont.)
POTENTIAL FAULTS AND POSSIBLE CAUSES / RESOLUTIONS