G. TROUBLESHOOTING
CONTROLLER FEEDBACK INDICATORS
Controller
AUDIBLE SIGNAL
Controller
VISUAL SIGNAL
Conditions
Action Required
1
1 Brief Beep at start
of dose.
Solid green LED for
duration of dose.
Normal: generating
aerosol mist, no errors
detected.
No action required.
MAGNAIR is on and
working properly.
2
1 Brief Beep at start
of dose.
Flashing orange-
green LED.
Low battery power.
Replace batteries or
use AC adapter.
3
1 Brief Beep at start
of dose followed by a
2-tone Beep.
Green LED followed
by flashing orange-
green and then
shuts off.
No connection
detected.
Two Beeps occurred
immediately after
starting.
No drug detected.
Two-tone Beep
occurred
10-30 seconds
after starting.
Check connection
between the Controller
and the Handset.
Check to confirm a
LONHALA vial has
been inserted and the
Handset clasp is
closed properly.
Tilt the Handset from
side to side then tap
the Handset to dislodge
LONHALA and restart
the Controller.
4
1 Brief Beep.
LED starts green,
then turns orange
and shuts off.
Treatment interrupted.
The Controller On/Off
button has been
intentionally or
unintentionally pressed.
Press the Controller
On/Off button to
resume treatment.
5
2 Brief Beeps.
LED flashes green
for 2 seconds, and
then shuts off.
Normal: end of dose.
No action required.
LONHALA vial contents
are empty.
6
2-tone Beep.
LED flashes
orange-green.
Maximum nebulization
time reached
(15 minutes).
Aerosol head clogged.
Press On/Off to
resume treatment
if medication is still
present. Clean the
Aerosol head.