Filling the nebuliser
• Insert the nebuliser in the holder on the compressor in-
tended for this purpose.
• If applicable, detach the inspiratory valve from the nebuliser.
• Pour the required quantity of
inhalation solution into the top
of the nebuliser.
Note the minimum and
maximum fill volumes
[see: General nebuliser data,
page 23]. If the nebuliser
contains too little or too much
liquid, the nebulisation and
consequently the therapy will
be less effective.
• Fit the inspiratory valve onto
the nebuliser.
If several inhalation solutions are to be used one after the
other:
• Rinse the nebuliser out with drinking water between the indi-
vidual applications.
• Shake excess water out of the nebuliser.
• Fill the nebuliser with the next inhalation solution as
described.