background image

   

   

 
Residual Volume 

Note  that  about  1mL  of  medication  cannot  be  nebulised  and  remains  unused  in  the 
medication  reservior;  this  residue  must  be  discarded.    This  in  intentional  and 

does 

not

 represent a malfunction.  The device switches off as soon as the membrane of 

the aerosol head is no longer coated with medication.  

 

 

 

 
 
 

 
eFlow® rapid easycare® cleaner   

 
Measure the time taken to nebulise precisely 2.5 ml isotonic saline 
solution 0.9% 

without inhaling

1.

 

If nebulising takes longer than 4 minutes, the aerosol head 
should be cleaned with the Easycare cleaning tool  

2.

 

If nebulising takes longer than 5 minutes the aerosol head should be replaced  

As a guide aerosol heads are efficient for 300-400 administrations therefore replace 
approximately: 

o

 

every YEAR if nebulising once per day  

o

 

every SIX months if nebulising twice per day (ask for 

a PARI 6-month timer)  

o

 

every THREE months if nebulising four times per day  

It is important to remember the easycare® head cleaner will not lengthen the life of 
the aerosol head and therefore should not be used any more than twice a week.   

How to use Video  

A video is available on our website under the product tab, eFlow rapid.  If you would 
like a copy of this video, please let us know. 

http://www.pulmomed.com.au/eFlowrapidvideo.html

 

 

 
Handling & Hygiene Brochure

                                                           

http://pulmomed.com.au/Brochures/PARI/Lower%20Airways/eFlow%20rapid/eFlow%
20Rapid%20cleaning.pdf

  

 

Last updated Monday 16 December 2013 

Reviews: