CLEANING THE NEBULISER

It is important that the nebuliser handset is cleaned immediately after each treatment to ensure proper performance, since 
viscous medications such as antibiotics may clog the nebuliser more easily than other medications. If you are unable to clean 
the nebuliser after every treatment, you must rinse the nebuliser to remove the medication residue. Then resume cleaning 
as soon as it is convenient.

1

2

Wash your hands as instructed on page 8. 

Please note that the aerosol head  
(i.e. the metal part) requires careful hand-
ling. The central metal parts should not be 
touched and the aerosol head should only 
be handled by the plastic collar.

Completely disassemble the nebuliser parts. 
Discard the filter pad safely after every 
treatment.

Remove the medication cap by twisting. Empty 
any remaining medication residue.

Rinse all parts of the nebuliser under a strong stream 
of warm water. 

WASHING: Mix a solution of warm water and 
washing-up liquid in a clean container. Wash 
all parts of the nebuliser in this solution. Rinse 
all parts thoroughly after washing and place 
between two clean towels or cloth to dry whilst 
keeping dust-free. Do not use kitchen towel. All 
parts must be completely dry before re-assembly. 

3

4

DISINFECTING THE NEBULISER

We recommend you disinfect the eFlow

®

rapid nebuliser handset every day.  

There are 2 options for disinfecting your nebuliser as below.

STEP 1:

Make sure the nebuliser has been 
cleaned prior to disinfecting (see  
page 14, “Cleaning the Nebuliser”)

STEP 2:

Boil all parts of the nebuliser in a clean 
pan for at least 5 minutes. The aerosol 
head must only be boiled with distilled 
water. 
Place on a clean towel to dry. Cover 
the parts with a clean towel to keep 
dust free. All parts must be dry before 
re-assembly.

1

2

STEP 1:

Make sure the nebuliser has been  
cleaned prior to disinfecting (see page 
14 “Cleaning your Nebuliser”)

STEP 2:

Place all parts of the nebuliser into a  
standard disinfector for baby bottles. 
Follow the instructions for use. 
Please note you should not use a micro-
wave disinfector for the aerosol head.
Place on a clean towel to dry. Cover the 
parts with a clean towel to keep dust free. 
All parts must be dry before re-assembly. 

OPTION 1 – BOILING 

OPTION 2 – STEAM DISINFECTION 

Summary of Contents for eFlow rapid

Page 1: ...A step by step guide for patients using the eFlow rapid eFlow rapid USER GUIDE FOR NEBULISED ANTIBIOTICS www pari com...

Page 2: ...I reusable nebulisers available for aerosol delivery of inhaled antibiotics but this booklet deals solely with the eFlow rapid This booklet acts as a supplement to not replacement for the Instructions...

Page 3: ...sol Delivery of Inhaled Antibiotics 6 Supplies Checklist 7 Handwashing Technique 8 Aerosol Therapy Set Up Nebuliser with exhalation filter 10 Aerosol Treatment 12 Cleaning the Nebuliser 14 Disinfectin...

Page 4: ...41G0500 Replacement Filter Pads Part No 041B0522 pack of 30 Part No 041B0523 pack of 100 PARI Filter Valve Set Replacement Filter Pads This is the eFlow rapid nebuliser system AEROSOL DELIVERY OF INHA...

Page 5: ...to handling supplies and medications Wet hands thoroughly with warm water Rinse hands thoroughly under running water Use antibacterial soap if possible Dry with a clean towel Thoroughly wash hands for...

Page 6: ...the nebuliser Replace the medication cap Read and understand the PARI nebuliser instructions for use 5 8 9 6 10 11 7 You should have 1 x Mouthpiece without valve or BABY mask for young children babies...

Page 7: ...enever possible ALTERNATIVELY if using the PARI BABY mask place the mask comfortably over the mouth and nose of the patient 1 4 2 5 6 3 Monitor for side effects If you experience any of the following...

Page 8: ...r Wash all parts of the nebuliser in this solution Rinse all parts thoroughly after washing and place between two clean towels or cloth to dry whilst keeping dust free Do not use kitchen towel All par...

Page 9: ...wear During its entire period of use an aerosol head should not be back washed more than 25 times with the easycare cleaning aid Backwash procedure Start the backwash by pressing the start button on t...

Page 10: ...an optional accessory that will minimise aerosol escaping into the room This may be important if you don t want family members or anyone other than the patient inhaling the antibiotic aerosol mist Ex...

Page 11: ...20 www pari com PARI Medical Ltd The Old Sorting Office Rosemount Avenue West Byfleet Surrey KT14 6LB E Mail infouk pari eu Phone 44 0 1932 341122 Fax 44 0 1932 341134...

Reviews: