NEBULISED ANTIBIOTICS
USER GUIDE
NEBULISED ANTIBIOTICS
USER GUIDE
How to assemble the easycare
Fill approx. 5 ml isotonic sodium chloride
solution (0.9 % saline) into the top opening
Connect the easycare cleaning aid to the eBase
®
controller. Attach the eFlow
®
rapid
nebuliser cable
to the hinge of the easycare reservoir and swivel
it downwards, so that the metal contacts engage.
Connect the other end of the cable to the eBase
®
controller.
Open the easycare cleaning aid. To open
press lightly on the sides and pull open.
Install the aerosol head and close the easy-
care reservoir. The eFlow
®
aerosol head must
be felt to snap into place upon insertion.
Please take care to ensure proper orientation.
1
3
2
4
Frequency of use
If needed, use the easycare no more
than 2 times per week after cleaning
the aerosol head.
i
Note
More frequent use is
not
beneficial,
and can lead to premature material
wear. During its entire period of use,
an aerosol head should not be back-
washed more than 25 times with the
easycare cleaning aid.
Backwash procedure:
Start the backwash by pressing the start
button on the eFlow
®
control unit. The pro-
cessing time for 5 ml is at least 10 min.
After rinsing, the eFlow
®
control unit auto-
matically switches off. If liquid is still present
in the funnel, please restart.
Important:
The easycare backwashing procedure is
performed as an additional measure to the
normal cleaning of the aerosol head.
EASYCARE CLEANING AID
• Measure how long it takes to
nebulise 2.5 ml isotonic sodium
chloride solution (0.9 % saline)
without inhaling.
• If the nebulisation time takes longer
than 4 minutes, it is recommended
to backwash the aerosol head with
the easycare cleaning aid.
Please ensure that the following steps are observed to optimise easycare backwashing:
1
2
3
Clean the nebuliser
parts and aerosol head
as described on page 14.
Use the easycare – for
detailed description see
page 17.
Disinfect the nebuliser parts
and aerosol head as described
on page 15.
Use with the eFlow
®
rapid