6

TECHNICAL DATA

6.1 General

Nominal energy

15.6 Wh

Nominal voltage

7.24 V

Nominal capacitance

2,150 mAh

Output extra-low voltage

12 VDC / 1.67 A

Housing dimensions

(W × H × D) 8.7 × 4.5 × 8.9 cm

Weight

160 g

Runtime

approx. 40 min. when fully
charged

Charging time

approx. 2 h until fully charged

Input

12 V DC / 1.67 A

Output

12 V DC / 1.67 A

6.2 Ambient conditions

Operation

Ambient temperature

+10 °C to +30 °C

Relative humidity

30% to 75% (non-condensing)

Atmospheric pressure

700 hPa to 1,060 hPa

Transportation and storage

Minimum ambient temperature (without
monitoring of relative humidity)

-25 °C

Maximum ambient temperature (with
relative humidity of up to 93%, non-
condensing)

+45 °C

Humidity

max. 93%

Atmospheric pressure

500 hPa – 1,060 hPa

– 27 –

Summary of Contents for 047

Page 1: ...PARI accumulator Type 047 2017 PARI GmbH Spezialisten f r effektive Inhalation 047D3011 C 08 17...

Page 2: ...de Gebrauchsanweisung 3 en Instructions for use 17...

Page 3: ...ETRIEBNAHME 7 Akku laden 7 Akku an Kompressor anschlie en 8 Akku von Kompressor trennen 9 WARTUNG 9 Reinigung des Akkus 9 Aufbewahrung 10 FEHLERSUCHE 11 TECHNISCHE DATEN 12 Allgemein 12 Umgebungsbedin...

Page 4: ...n Sicherheitshinweisen Sicherheitsrelevante Warnungen sind in dieser Gebrauchs anweisung in Gefahrenstufen eingeteilt Mit dem Signalwort WARNUNG sind Gefahren gekenn zeichnet die ohne Vorsichtsma nahm...

Page 5: ...PARI Die Verwendung von Fremdprodukten kann den Akku besch digen Der Akku darf nicht zerlegt werden Entsorgen Sie ihn entspre chend den landesspezifischen Entsorgungsregeln 2 PRODUKTBESCHREIBUNG 2 1...

Page 6: ...dann noch ca 10 Min d h der Akku muss bald aufgeladen werden Wenn alle vier LEDs blinken hat der Akku eine Fehlfunktion siehe FEHLERSUCHE Seite 11 2 4 Produktkombinationen Der Akku ist geeignet f r d...

Page 7: ...Anwendung das Geh use des Akkus auf Unversehrtheit Der Akku darf nicht geladen oder verwen det werden wenn er besch digt ist Die Akkuladung des Neuproduktes betr gt bis zu 60 Laden Sie vor dem ersten...

Page 8: ...von der Netzsteckdose und dann vom Akku Info Der Akku kann auch ber den 12 V Autoadapter geladen werden Ist der Kompressor aus geschaltet aber ber den 12 V Autoadapter an einer 12 V Stromversorgung a...

Page 9: ...r durch Stromschlag Fl ssigkeiten k nnen Strom leiten wodurch die Gefahr ei nes Stromschlags entsteht Schalten Sie vor jeder Reinigung den Kompressor aus und ziehen Sie das Netzteil bzw den 12 V Autoa...

Page 10: ...takt Seite 16 4 2 Aufbewahrung Trennen Sie das Netzteil bzw den 12 V Autoadapter vom Akku Lagern Sie den Akku gesch tzt vor W rmequellen jeglicher Art z B vor anhaltender direkter Sonneneinstrahlung L...

Page 11: ...Es liegt eine St rung am Kompressor oder am Akku vor Setzen Sie den Akku zur ck Schalten Sie den Kompressor aus Trennen Sie den Akku vom Kompres sor siehe Akku von Kompressor trennen Seite 9 Warten Si...

Page 12: ...ung Eingang 12 V 1 67 A Ausgang 12 V 1 67 A 6 2 Umgebungsbedingungen Betrieb Umgebungstemperatur 10 C bis 30 C Relative Luftfeuchtigkeit 30 bis 75 nicht kondensierend Luftdruck 700 hPa bis 1 060 hPa T...

Page 13: ...kku nach Wahl des Herstellers repariert ausgetauscht oder es wird der Einkaufspreis zur ckerstattet Der Austausch des Akkus kann sowohl in das gleiche als auch in ein mindestens ver gleichbares Modell...

Page 14: ...Abnutzung ausgesetzt sind Im Falle einer Reklamation kontaktieren Sie unbedingt zu n chst Ihren Fachh ndler oder die PARI GmbH F r den Versand von Lithium Ionen Akkus gibt es spezielle Sicherheits vo...

Page 15: ...ukt nicht ber den Hausm ll entsorgt werden Es sind die jeweiligen landesspezifischen Entsorgungsregeln zu beachten z B Entsorgung ber die Kommunen oder H ndler Hersteller E24 ECE Genehmigung f r die I...

Page 16: ...3479 Ellwangen Vertrieb PARI GmbH Moosstr 3 82319 Starnberg 7 4 Kontakt F r Produktinformationen jeder Art im Fehlerfall oder bei Fra gen zur Handhabung wenden Sie sich bitte an unser Service Center T...

Page 17: ...ble battery 21 Connecting the rechargeable battery to the compressor 23 Disconnecting the rechargeable battery from the compressor 23 MAINTENANCE 24 Cleaning the rechargeable battery 24 Storage 25 TRO...

Page 18: ...1 3 Structure of safety instructions Safety critical warnings are categorised according to hazard levels in these instructions for use The signal word WARNING is used to indicate hazards which withou...

Page 19: ...e to open fire The rechargeable battery is not splashproof Protect it from moisture and liquids Use only the PARI power adapter or 12 V car adapter supplied with the compressor to charge the rechargea...

Page 20: ...evel when charging as follows Charging process display goes out when rechargeable battery is fully charged Charge level 75 and higher Charge level 50 to 74 Charge level 25 to 49 The LEDs indicate the...

Page 21: ...posing a risk of electric shock Never take hold of the power adapter with damp or wet hands The rechargeable battery can be charged using either the power adapter supplied with a type 047 compressor...

Page 22: ...battery is being charged The charging process display goes out as soon as the rechargeable battery is fully charged see Explanation of the LEDs on the rechargeable battery page 20 When the charging p...

Page 23: ...two components together until the engaging hook on the re lease button engages in the bottom of the battery 3 3 Disconnecting the rechargeable battery from the compressor Disconnect the power adapter...

Page 24: ...quids get into the interior of the rechargeable battery this may cause a fault Do not spray any liquid onto the rechargeable battery Disconnect the battery from the com pressor see Disconnecting the r...

Page 25: ...tained for as long as pos sible store the charged battery at room temperature ap prox 21 C If the battery is not going to be used for a long period it must be charged again after 6 months because othe...

Page 26: ...he compressor or the rechargeable battery Reset the rechargeable battery Switch the compressor off Disconnect the battery from the compressor see Disconnecting the rechargeable bat tery from the com p...

Page 27: ...2 h until fully charged Input 12 V DC 1 67 A Output 12 V DC 1 67 A 6 2 Ambient conditions Operation Ambient temperature 10 C to 30 C Relative humidity 30 to 75 non condensing Atmospheric pressure 700...

Page 28: ...on the rechargeable battery will be repaired or replaced or the purchase price will be refunded If it is replaced the re placement battery will either be the same model or a model that is at least com...

Page 29: ...omplaint you must first contact your specialist dealer or PARI GmbH Special safety regulations apply for sending lithium ion rechargeable batteries and these must be complied with The warranty certifi...

Page 30: ...ities or dealers Manufacturer E24 ECE approval for operation in motor vehicles This product satisfies the requirements of 2004 108 EC EMC directive 2006 66 EC Battery directive and 2011 65 EU RoHS dir...

Page 31: ...PARI accumulator Type 047 CERTIFICATE OF GUARANTEE We guarantee this accumulator for 6 months commencing on the date of purchase...

Page 32: ...utor PARI GmbH Spezialisten f r effektive Inhalation Moosstra e 3 82319 Starnberg GERMANY Tel 49 0 8151 279 0 Fax 49 0 8151 279 101 info pari de www pari com Bar code label Application no Confirmation...

Reviews: