5
TROUBLESHOOTING
In the event of faults that are not listed in this chapter, or if the
suggested procedure does not correct the fault, contact the
PARI GmbH Service Center.
Fault
Possible cause
Procedure
The
rechargeable
battery is
plugged into
the
compressor,
but the
compressor is
not working.
The battery is flat.
Connect the power
adapter to the
rechargeable battery to
charge it.
There is a fault in the
compressor or the
rechargeable battery.
Reset the rechargeable
battery:
•
Switch the compressor
off.
•
Disconnect the battery
from the compressor
[see: Disconnecting
the rechargeable bat-
tery from the com-
pressor, page 23].
•
Wait at least
10 minutes before op-
erating the re-
chargeable battery
again.
•
If the problem persists,
try charging the bat-
tery [see: Charging the
rechargeable battery,
page 21].
•
If the problem still can-
not be corrected, con-
tact Technical Service
[see: Contact,
page 30].
All 4 LEDs are
flashing.
– The battery is over-
loaded (> 3.5 A).
– The battery is over-
heated.
– The battery is flat
(deeply dis-
charged).
– 26 –