DEVICE MAINTENANCE AND CLEANING
TRAINING
7
©
Parata
2014
ATP Device Components
1.
Medication Drawer.
Drawer that
houses the canisters that contain the
medication. The
ATP Device
has 4 rows
of drawers stacked 5 or 6 high, with a
maximum of 24 medication drawers.
The medication drawers are always
locked for safety. Always pull out 1
drawer for 1 row when opening. Pulling
out more than one drawer in a column simultaneously is not possible.
Note: For safety purposes, if multiple drawers are open at the same time, the machine
could potentially tip.
2.
Control Panel.
Packaging status and alarms are displayed on this panel. You will not
need to use these buttons, as all commands are initiated by your software application
on the screen. Only use these buttons if instructed to do so by
Parata
Support.
3.
Package Exit.
The packets are advanced to this exit though the internal conveyor.
4.
Front Door.
The door protects the operator from the lower packaging unit (motor,
heater, etc.). The device will not operate when the front door is open.
5.
Lower Packaging Unit.
The mechanism for making a packet including the conveyor,
shutter, sealer, cutter, and printer.