Daily maintenance
When removing stains from parts
Perform steps 1 to 5 below.
1.
Wipe off stains with a cloth soaked in a neutral detergent diluted with water and wrung out well.
2.
Wipe off any residual detergent with a cloth soaked in water and wrung out well.
3.
Wipe away any leftover moisture with a dry cloth.
4.
At the conclusion of cleaning, visually inspect all surfaces of the oxygen cylinder holder.
5.
If not visually clean, repeat cleaning the cleaning procedure.
If the oxygen cylinder holder cannot be visually cleaned, the oxygen cylinder holder
should be discarded.
When disinfecting parts
Use one of the disinfectants listed below at the specified concentrations by following the instructions in the
instruction manual for the disinfectant.
76.9% - 81.4% Ethanol for disinfection
0.05% - 0.2% Benzalkonium chloride (e.g., Osvan)
0.05% - 0.2% Benzethonium chloride (e.g., Hyamine)
0.05% Chlorhexidine gluconate (e.g., Hibitane)
0.02% - 1.0% Sodium hypochlorite (e.g., Milton)
* We recommend to disinfect the oxygen cylinder holder before use by another patient.
●
Do not use volatile cleaning solutions (such as paint thinner, benzine, or
gasoline) or cresol. Discoloration or deterioration may result.
●
Do not use disinfectants other than those listed above. Damage or
deterioration may result.
●
Thoroughly wipe off parts with a damp cloth after using a neutral detergent.
Not doing so may cause cracking of the resin.
●
If the oxygen cylinder holder is stained, wipe off promptly. Leaving it as is may
cause permanent stains.
●
An ozone sterilizer, autoclave, bed cleaning machine, etc. cannot be used for the
oxygen cylinder holder.
●
Do not sterilize. The oxygen cylinder holder is not intended for sterilization and has
not been tested for compatibility with healthcare sterilization systems including
ozone, moist heat/steam, dry heat, ethylene oxide vaporized Hydrogen peroxide
and other chemical sterilizers. Exposure of the oxygen cylinder holder to these
processes may damage the oxygen cylinder holder and render it non-functional.
Long-term storage
Avoid a location subject to high temperature or humidity, low temperature, dryness or excessive
dust. Store the oxygen cylinder holder at a temperature between -
10˚C and 50˚C, and humidity
between 30% RH and 90% RH.
Keep the instruction manual in a safe place so that it does not become lost.
Disposal
When disposing of the product, observe the local rules and regulations.