background image

- 2 - 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Stock Code:  BAC-AF-0268-00 

 
 
 
 

THE AIRSCAN MOBILE CO MONITOR IS NOT 

FOR USE IN EXPLOSIVE AREAS

 
 

3.0

 

INSTALLATION 

 
3.1

 

The PanBlast™ AirScan Mobile CO Monitor is 
designed to operate in a portable or fixed 
application. In all applications it is important to 
consider: 

 

-

 

Availability of electrical power – 220VAC 
or 110VAC or 12VDC. 

-

 

Keeping volume of sample line small. 

-

 

User ability to hear alarms. 

-

 

Instrument and sample line stability and 
protection from damage. 

-

 

Instrument thermal environment between 

4.5

ºC 

(40

º

F) and 82

ºC 

(180

º

F). 

-

 

Access to controls for operational checks 
and calibration. 

-

 

Sample line originates downstream of 
filtering and air treatment. 

 

3.2

 

The sample air from the breathing air system 
should be set up to avoid any undue moisture 
problems. Good practice and use of water legs 
and vertical sampling from lines should be 
observed. Refer to Figure 2 for optional ways to 
limit any condensation problems. 

 

3.3

 

The sample line should be provided with the 
necessary reduction plumbing to fit the 6mm 
(1/4") NPT female opening on the AirScan 
Mobile CO Monitor. It is recommended that the 
ID of the sample line be as small as possible 
down to 1.5mm (1/16”) ID. 

 

3.4

 

The electrical connection is set up as standard 
220 VAC 50Hz. Ready shut-off of fixed 
installations should be provided. Sample line 
volume is critical; keep line as short as possible. 
3mm (1/8") OD x 1.5mm (1/16") ID 
Polyurethane tubing is commonly used as the 
sample line although it may require physical 
protection in exposed areas. Galvanized and 
PVC material should not be used.

 

 

 
 
 
 

 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Figure 2 

 

4.0

 

OPERATION 

 
4.1

 

The PanBlast™ AirScan Mobile CO Monitor 
operation is based on taking a small sample 
from the breathing air system concurrent with 
the air being delivered to personnel in 
approved breathing apparatus. This small 
sample is exposed to a sensor capable of 
detecting low levels of carbon monoxide (CO) 
and other toxic gases or vapors. Using known 
levels of carbon monoxide gas (10 ppm CO for 
example) it is possible to calibrate the AirScan 
Mobile CO Monitor to generate alarms when the 
sensor is exposed to levels equal to or higher 
than the level of the calibration gas.  

 

Since it is critical that the sample air be 
continuously monitored the PanBlast

TM

 AirScan 

Mobile CO Monitor also has a pressure sensor 
that confirms the sample flow. If the sample 
airflow is too low or too high, the instrument 
will initiate a flow alarm. (Refer to the section of 
the manual on visual annunciators, controls, 
and audio alarms for more information on these 
alarms.) 
 

DANGER 

Line Cord 

Tee in piping 
or access point 
in filtration 
system 

To 

Respirators 

From 

Filtration 

 

Sample Tube 

Drip leg used 
in fixed 
applications 

Drain 

Summary of Contents for ZVP-PC-0080-00

Page 1: ...PRODUCT OPERATING MANUAL PANBLASTTM AIRSCAN MOBILE CO MONITOR Manual Number ZVP PC 0070 00...

Page 2: ...SECTION 1 GENERAL INFORMATION 2 INTRODUCTION 3 INSTALLATION 4 OPERATION 5 INITIAL SET UP 6 RESPONSE CHECK 7 OPERATING MODE 8 CALIBRATION 9 TROUBLESHOOTING GUIDE 10 PARTS LISTING...

Page 3: ...ectrical trades personnel 1 3 4 Under no circumstances should any safety interlocks lockouts or features be altered or disabled in any way 1 3 5 All equipment must be isolated from the compressed air...

Page 4: ...t off of fixed installations should be provided Sample line volume is critical keep line as short as possible 3mm 1 8 OD x 1 5mm 1 16 ID Polyurethane tubing is commonly used as the sample line althoug...

Page 5: ...put port provides the capability to introduce calibration gases to the instrument sensor when the calibration Run valve is in the calibration mode 8 Faceplate door screw 9 Internal sample flow regulat...

Page 6: ...assembly to the closed position and remove the test gas from the instrument test gas port 6 Move the RUN CAL switch to the RUN position 6 4 Check and confirm that the flow rate returns to the GREEN co...

Page 7: ...edle knob of the calibration valve assembly to open Monitor the flow bar graph display to ensure GREEN condition and allow the impurity free gas to flow for at least one minute Check the air quality i...

Page 8: ...board has two 2 12VDC output Molex connectors J3 and J5 shown in Figure 5 One is used to power the internal alarm horn the other is unused and can power an external low current alarm device Maximum cu...

Page 9: ...sor loosen the screw 8 Figure 3 and open the faceplate When the enclosure is open follow the directions listed below 1 Carefully remove the 2 screws from the sensor block and lift the block off the se...

Page 10: ...Contact manufacturer for service Alarm horn will not sound Faulty or unplugged horn Plug horn in Replace horn Bad air Air quality indicator is RED Check intake for contamination Out of calibration Re...

Page 11: ...t Parts Stock Code Description Qty BAC AF 0270 00 CO Test Gas 1 BAC AF 0271 00 Impurity Free Air Test Gas 1 BAC AF 0269 00 Calibration Valve Assembly 1 BAC AF 0341 00 AirScan Sensor 1 BAC AF 0342 00 F...

Reviews: