1. Plug the AC cord directly into the AC outlet. Do not use an
isolation transformer for this check.
2. Connect “A” to exposed metallic part on the set. And
connect “B” to a good earth ground, as shown in .
3. Use an AC voltmeter, with 1 k
W
/V or more sensitivity, to
measure the potential across the resistor.
4. Check each exposed metallic part, and measure the
voltage at each point.
5. Reverse the AC plug in the AC outlet and repeat each of the
above measurements.
6. The potential at any point should not exceed 0.25 V RMS.
A leakage current tester (Simpson Model 229 or equivalent)
may be used to make the hot checks, leakage current must
not exceed 1/2 mA. In case a measurement is outside of
the limits specified, there is a possibility of a shock hazard,
and the equipment should be repaired and rechecked
before it is returned to the customer.
Figure 1
1 Safety Precaution
1.1. General Guidelines
1. IMPORTANT SAFETY NOTICE
There are special components used in this equipment which are important for safety. These parts are marked by in the
Schematic Diagrams, Circuit Board Layout, Exploded Views and Replacement Parts List. It is essential that these critical parts
should be replaced with manufacturer’s specified parts to prevent X-RADIATION, shock fire, or other hazards. Do not modify
the original design without permission of manufacturer.
2. An Isolation Transformer should always be used during the servicing of AC Adaptor whose chassis is not isolated from the AC
power line. Use a transformer of adequate power rating as this protects the technician from accidents resulting in personal injury
from electrical shocks. It will also protect AC Adaptor from being damaged by accidental shorting that may occur during
servicing.
3. When servicing, observe the original lead dress. It a short circuit is found, replace all parts which have been overheated or
damaged by the short circuit.
4. After servicing, see to it that all the protective devices such as insulation barriers, insulation papers shields are properly
installed.
5. After servicing, make the following leakage current checks to prevent the customer from being exposed to shock hazards.
1.1.1. Leakage Current Cold Check
1. Unplug the AC cord and connect a jumper between the two prongs on the plug.
2. Measure the resistance value, with an ohmmeter, between the jumpered AC plug and each exposed metallic cabinet part on
the equipment such as screwheads, connectors, control shafts, etc. When the exposed metallic part has a return path to the
chassis, the reading should be between 1M
W
and 5.2M
W
. When the exposed metal does not have a return path to the chassis,
the reading must be infinity.
1.1.2. Leakage current hot check (See Figure 1)
3
SDR-H80P
Summary of Contents for SDR-H90P
Page 8: ...2 5 How to Recycle the Lithium Battery U S Only 8 SDR H80P...
Page 12: ...3 2 2 Precautions for installing HDD 12 SDR H80P...
Page 16: ...4 Specifications 16 SDR H80P...
Page 18: ...18 SDR H80P...
Page 21: ...Fig 1 3 21 SDR H80P...
Page 22: ...Fig 1 4 22 SDR H80P...
Page 23: ...Fig 1 5 23 SDR H80P...
Page 24: ...Fig 1 6 Fig 1 7 To exit the Service Menu Unplug the AC Cord 24 SDR H80P...
Page 26: ...7 Service Fixture Tools 7 1 Service Tools and Equipment 26 SDR H80P...
Page 30: ...Fig D7 Fig D8 Fig D9 Fig D10 30 SDR H80P...
Page 31: ...Fig D11 Fig D12 Fig D13 Fig D14 31 SDR H80P...
Page 32: ...Fig D15 32 SDR H80P...
Page 36: ...9 3 1 Main P C B 9 3 Location for Connectors of the Main P C B and Sub P C B 36 SDR H80P...
Page 37: ...9 3 2 Sub P C B 37 SDR H80P...
Page 55: ...13 Parts and Exploded Views 13 1 Exploded Views 13 1 1 Frame Casing Section 55 SDR H80P...
Page 56: ...13 1 2 LCD Section 56 SDR H80P...
Page 57: ...13 1 3 Camera Lens Section 57 SDR H80P...
Page 58: ...13 1 4 Packing Parts Accessories Section 58 SDR H80P...